Administration Console
Uploading a Custom Calibration File
You can upload your custom calibration settings file to the
RTSA7550
via the
administrative console.
Note: You must restart the
RTSA7550
for the new calibration settings to take effect.
Note:
Custom calibrated values might get over-written by a new firmware update. After
restarting the unit once the firmware update is done, check the Calibration tab of the
administration console to see which calibration file source is listed on the top of the
Filename list (ie. “System” or your customized file).
1.
Click the "Calibration" link in the left column menu of the administrative console
web page. The "
RTSA7550
Device Calibration" page will appear.
2.
Click “Browse”. An Explorer window will open on your PC.
3.
Navigate to the location of the desired file and click “Open”. The selected file
name appears beside the “Browse” button in the Administration console.
4.
Click “Upload”. When successful, the custom calibration file is uploaded to the
RTSA7550
and appears in the list of calibration files. The most recent uploaded
file
will be listed at the top of the table.
5.
Restart the
RTSA7550
by using the administrative console or toggling the power
switch. The boot up will take longer than usual after the restart as the embedded
system applies the new calibrated values.
23
BNC
RTSA7550
Real-Time Spectrum Analyzer User Guide