7.2 Basic Operation
The following steps
must
be carried out before communication with the 525 may take place:
Ensure that the proper drivers have already been installed on the remote computer. These drivers will
need to be installed for the remote computer to interface to the 525
. Reference the “525 Setup
USB”
section for more information on installing the proper drivers.
Open the 525 application by double clicking on the application which can be found on the included CD.
Once the 525 application recognizes the COM Port the 525 is attached to, a connection will be
established, and communication may now be carried out.
7.2.1
System Section
The System Section of the 525 application only affects the 525
‟s system parameters. As shown below, the
following system parameters may be altered:
Figure 7-2
–525 Application: System Parameters
Pulse Mode:
Changes the system‟s output mode to Continuous, Single Shot, Burst, or Duty Cycle.
Period:
Sets the System‟s T
0
period. Valid inputs are 200ns
–1000s set in 10ns increments.
System Resolution:
Changes the system resolution between 4ns and 5ns. This affects the period,
widths and delays. This will also affect some of the system performance specifications.