background image

1.Brief Introduction  

  * It meets the standards of both ISO2178 and ISO-2361 as well as DIN, ASTM and 
BS. Suitable for the laboratory and for use in harsh field conditions. 
* The gauge can  measure the thickness of non-magnetic coating layers on magnetic 
substrate.Such as: (aluminum, chrome, copper, enamel, rubber, paint) on magnetic 
substrate (steel, iron, alloy and magnetic  stainless steel).. 
* The gauge  can measure the thickness of non-metals coatings layers covered on 
non-magnetic metals substrate. Such as: (enamel, rubber, paint, vanish, plastic 
anodic-oxide layer) covered on non-nagnetic metals substrate (aluminum, brass, zinc 
and nonmagnetic stainless steel). 

2.Characteristic 

* Small style design, conveniently carry    
* Complete intelligentized design, easily operate 
* Rapid measure, excellent repeated performance    
* Zero adjustment, second point adjustment 
* LCD digital display  
* The metric system /English system unit transform  
* There is bee ring song prompt during operation   
* Two types of probe heads design, which is applicable to different test situations 

3.Display Instruction  

μ

m indicate value unit 

μ

m=1/1000mm

 

* CAL O  ZERO Adjustment  
* Fe  Magnetism substrate measure state  
* NFe Non-magnetism substrate 
   Measure state 
* 9999  restore factory setting 
 

 

4.Technology Parameter  

Range 

   : 

0-1000

um

 

Accuracy 

   : 

±[(

1%

3%

)]H±

1.5

um

 

* Resolution   

 

 

: 0.1

um

 (0~99.9

um

                          

     

 

 

  1um (over 100um) 

* Environment temperature 

: 0-40

°C

 

* Relative humidity   

 

:

 ≤

85% 

* Power supply 

 

 

: 1.5V×2 

* Weight 

 

 

 

: 98g  

Dimension 

   : 

102×66×24mm 

4.1 Standard Delivery 

Main unit(1000um)  1pc 
Instruction manual  1pc 
Substrate  

2pc 

Carrying case 

1pc 

Measuring foil set  1set 

 

Operation 

ETA-0832 

Manual 

Rev:00                                                               

  3  / 6

 

 

Summary of Contents for ETA-0832

Page 1: ...S METER BMS Bulut Makina Sanayi Ve Ticaret Ltd Şti İkitelli Organize Sanayi Bölgesi Dolapdere Sanayi Sitesi Ada 4 No 7 9 İkitelli İSTANBUL TÜRKİYE Phone 90 212 671 02 24 671 02 25 Fax 90 212 671 02 26 web www bulutmak com e mail bms bulutmak com ...

Page 2: ...tic 3 3 Display Instruction 3 4 Technology Parameter 3 4 1 Standard Delivery 3 5 Measuring Procedure 5 6 Calibration 5 7 Battery Replacement 5 8 Considerations 6 9 Restore Factory Setting 6 10 Notes 6 Operation ETA 0832 Manual Rev 00 2 6 ...

Page 3: ...carry Complete intelligentized design easily operate Rapid measure excellent repeated performance Zero adjustment second point adjustment LCD digital display The metric system English system unit transform There is bee ring song prompt during operation Two types of probe heads design which is applicable to different test situations 3 Display Instruction μm indicate value unit 1 μm 1 1000mm CAL O Z...

Page 4: ... MEASURING FOIL DIGITAL SCREEN MAIN UNIT SUBSSTRATE FE PROBE CARRYING CASE Figure 1 ON OFF KEY VALUE INCREASING KEY ZEROING KEY MICRON MICROINCH UNIT TRANSFORM KEY VALUE DECREASING KEY Operation ETA 0832 Manual Rev 00 4 6 ...

Page 5: ...tched off by pressing the POWER key 5 7 Pressing the um mil key can change the measurement unit um or mil 6 Calibration 6 1 Zero adjustment Zero calibration for Fe and NFe should be carried out separately Take the iron substrate if Fe on the Display and take the aluminium substrate if NFe on the Display Place the probe on the substrate steadily Press the ZERO key and CAL0 OK will be on the Display...

Page 6: ...ER key at the same time and then release the POWER key firstly 2 seconds later release the um mil key 9 2 2 When the 9999 appears on the display press the POWER key the factory setting is restored 10 Notes It is strongly recommened that no changes should be made to the value of Ln Generally the larger the value of Ln the smaller the reading on a same thickness A little variation of value of Ln wil...