7
Figure 9 Alarm menu
4.3. Use
4.3.1
Installation of indenter: make the indenter into the hole axis, rotating set screws (2), soft
reduction in pressure to the axis of the flat head position, and then try to install the lift station
directly on the screw, and then try Sample firmly placed on the test platform to ensure the process
does not occur in the test and flexural displacement, rotating roller (4) to test Taiwan rose slowly,
the specimen and pressure head gently touch, rotating set screws tightly, turn the roller, pressure
head and try to block out.
Figure 10 structure
4.3.2
Turn on the power switch, the machine itself to initialize, display the initial interface, a large
lever to enter the work automatically adjust the starting position. Startup force value set in 612.9N
(62.5Kgf) position, the boot time set to 15 seconds, if you want to choose other security forces and
load time testing, please refer to the Features control panel.
4.3.3
After preparation ,Put test piece on stage, Rotate hand wheel , Work piece contact pressure
to be the first test force while also beginning to show, when the test force must be close to
automatic loading values rose slowly to reach the automatic loading value, the device will issue a
ALARM
Press <OK> CHECK
1
2
3
4
1.
upper cover
2.
elevating screw
3.
switch
4.
nut
5.
turntable
6.
eyepiece
5
6