bmcm AP8a Manual Download Page 4

AP8a 

 

 

Seite 4 

3

 

Connection examples for using the AP8a with 5B amplifiers 

The most different 5B amplifiers can be used with the AP8a. Please see the relevant data sheets of the 
5B modules for further connection examples. 

3.1

 

MA-U: Voltage measurement (DC and AC decoupled) 

The input is differential (balanced). 
Close jumpers J5 and J7 of the MA-U for AC decoupling to suppress 
DC components of the signal. AC decoupling can only be used in the 

0.5V and 

1V measuring range.  

3.2

 

MA-UNI: Strain gage measurement with DC 

Strain gages are resistors operated in bridge circuits. The EX voltage is 

2.5V DC. The input amplifier is operated in differential mode. If 

necessary, the sensor lines compensate for line losses.  
In the case of bridge extensions, precise supplementary resistors must 
be used (0.1%; TC15). If using 100

 bridges, it is only possible to 

supply with +2.5V, dividing the measuring range in half. 
The Jumpers J12, J13 of the MA-UNI must only be closed in 6-wire 
applications. For some applications, it is better to connect the screen to 
earth and not to the internal ground (0EX).  

4

 

Important notes for using the AP8a 

 

The AP8a is only suitable for extra-low voltages 

 please observe the relevant regulations! 

 

An electrically isolated power unit (with CE) must be used for power supply. 

 

All  accessible  pins  are  electrostatic  sensitive  devices.  Provide  an  earthed  conductive  work  place  when 
installing. 

 

Only use non-solvent detergents for cleaning. The product is designed to be maintenance-free.  

 

If using the direct switches, the relevant LO input is connected with the succeeding DAQ system and therefore 
may be connected to ground. Open inputs should be closed. 

 

The board must not be used for safety-relevant tasks. With the use of the product, the customer becomes 
manufacturer by law and is therefore fully responsible for the proper installation and use of the product. In the 
case  of  improper  use  and/or  unauthorized  interference,  our  warranty  ceases  and  any  warranty  claim  is 
excluded. 

 

Do not dispose of the product in the domestic waste or at any waste collection places. It has to be either duly 
disposed according to the WEEE directive or can be returned to bmcm at your own expense. 

5

 

Technical data (typical at 20°C, after 5min., 9-40V supply) 

  Electrical data 

Power supply: 

9..40V (max. 1A DC, default setting ex works) or +5V DC ±5% (max. 1.5A DC) 

Max. current to be drawn for the modules: 

max. 1A (

 total of current of all individual modules) 

Electrical isolation: 

depending on the module in use; in case of direct switching no electrical isolation! 

Max. permissible potentials: 

60V DC acc. to VDE, max. 1kV ESD on open lines 

  General data 

Temperature ranges // Relative humidity: 

operating temp. 0..70 C, storage temp. -25..70 C // 0-90% (not condensing) 

CE standards: 

EN61000-6-1, EN61000-6-3, EN61010-1; for decl. of conformity (PDF) visi

www.bmcm.de

 

ElektroG // ear registration: 

RoHS and WEEE compliant // WEEE Reg.-No. DE75472248 

Dimensions (L x W x H): 

backplane: 18cm x 10cm x 2cm; with DIN rail carrier and bracket: 18.3cm x 10.5cm x 4.3cm 

Available accessories (optional): 

7-pin pluggable spring clamp terminals ZU7ST (package of 8), power supply ZU-PW40W,  

waterproof housing ZU-PBOX-PG, ZU-PBOX-LAN 

Warranty: 

2 years from date of purchase at bmcm, claims for damages resulting from improper use excluded 

Manufacturer: BMC Messsysteme GmbH. Subject to change due to technical improvements. Errors and printing errors excepted. Rev. 4.1  

02/05/2020

 