bmcm AMS42-LAN16fx Manual Download Page 7

AMS42-LAN16fx 

Seite 7 

3.6

 

Digital Channels and Counters 

The AMS42-LAN16fx features two digital ports with 16 inputs or outputs each. The lines are bidirecti-
onal and set in groups of eight. Additionally integrated are three counters (quadrature decoders) for 
pulse, frequency, position, or pulse time measurement.  
With the option 

AMS-DIG8

 (see chapter 6.4), 8 inputs and outputs each can be used as galvanically 

isolated optocoupler inputs and relay outputs available at the back of the device. 

3.6.1

 

Digital Assignment 

All lines are led to the 37-pin D-Sub female connector at the device front. The pin assignment is as 
follows: 

D-Sub 37
(Digital In/Out) 

Assignment 
(Port/Bit) 

D-Sub 37
(Digital In/Out) 

Assignment 
(Port/Bit) 

11 

A/1 

B/1 

30 

A/2 

20 

B/2 

12 

A/3 

B/3 

31 

A/4 

21 

B/4 

13 

A/5 

B/5 

32 

A/6 

22 

B/6 

14 

A/7 

B/7 

33 

A/8 

23 

B/8 

15 

A/9 

B/9 

34 

A/10 

24 

B/10 

16 

A/11 

B/11 

35 

A/12 

25 

B/12 

17 

A/13 

B/13 

36 

A/14 

26 

B/14 

18 

A/15 

B/15 

37 

A/16 

27 

B/16 

10, 19, 29 

DGND 

9, 28 

5V, max. 10mA 

The digital inputs and outputs are protected by serial resistors. Input voltages beyond the permitted voltage range of

0V..5V can damage the device.

By default, port A is set to input, port B to output.

Digital inputs are always read synchronously in time together with the analog inputs.

Make sure the digital outputs of the device do not drive against the outputs of your signals.

3.6.2

 

5V Auxiliary Voltage 

The AMS42-LAN16fx provides an auxiliary voltage (e.g. for sensor supply) at pin 9, 28 of the D-Sub37 
female. The 5V DC output (max. 10mA) is protected by a fuse (multifuse). In case of overload, it is suf-
ficient to interrupt the power supply (turn off PC or disconnect the consumer load). After app. 1 min., 
the multifuse will be regenerated. 

Via Paolo Uccello 4  -  20148 Milano   

Tel +39 02 48 009 757   Fax +39 02 48 002 070      [email protected]   www.dspmindustria.it

Summary of Contents for AMS42-LAN16fx

Page 1: ...2 Analog OUT Signals are connected at 8 amplifier inputs with 16 bit resolution and 250kHz total sampling rate Ad ditional 8 analog inputs and 2 analog outputs are available directly and without ampl...

Page 2: ...ww bmcm de ams42 lan16fx en in the tab Downloads in the category LAN Admin Configuration Software 3a Step Measuring with NextView To be able to use your measurement system with NextView you just need...

Page 3: ...address this IP address must have been given out only once Each IP address contains the network number identifying the net your PC belongs to Please make sure the PC and the AMS42 LAN16fx are in the s...

Page 4: ...DIN coupling for 12 40V power supply use a supply cable with a cross sectional area 1mm2 and turn on the AMS42 LAN16fx by pressing the switch to 1 ON see chapter 3 1 b Connect the AMS device to the n...

Page 5: ...supply unit ZU PW40W 24V 1 67A DC is optionally available Pin Assignment 1 n c 2 ground GND 3 12 40V DC Connect the shield of the power supply with earth PE 3 4 Network The PC data acquisition server...

Page 6: ...ntial deviations however should not exceed 60V DC ac cording to VDE Association of German Electrical Engineers 3 5 2 Analog Outputs and Other Analog Inputs The 37 pin D Sub female connector at the bac...

Page 7: ...23 B 8 15 A 9 5 B 9 34 A 10 24 B 10 16 A 11 6 B 11 35 A 12 25 B 12 17 A 13 7 B 13 36 A 14 26 B 14 18 A 15 8 B 15 37 A 16 27 B 16 10 19 29 DGND 9 28 5V max 10mA The digital inputs and outputs are prote...

Page 8: ...sed to connect an incremental encoder consi ders the counting direction by decoding a second phase shifted signal To activate the counting function the respective digital lines of the AMS42 LAN16fx mu...

Page 9: ...e pin assignment corresponds to the 5B modules of Analog De vices BURR BROWN etc However an additional 0EX pin has been introduced which is particularly suitable for ungrounded shiel ding This is a sp...

Page 10: ...2 15 power supply modules 9V 12V 15V 6 2 Plug in Cassettes AMS K Line The amplifier system can individually be equipped with 8 plug in cas settes or 16 with option AMS EXT8 see chapter 0 Each cassette...

Page 11: ...cassettes a connection between power supply ground and chassis of the device is made via the module screw The Gain is adjusted to even values Therefore only 64000 with 16 bit values of the full range...

Page 12: ...er BNC Thermo on the front panels of the cassettes other analog channels at a 37 pin D Sub female at the device back Digital connections all channels at a 37 pin D Sub female at the device front LAN c...