background image

OratectXP

®

 Reference Guide

 

 

               

*OratectXP

®

 is for forensic use only

 

 

 

                            

Rev: 

B, 

 

 

 

 

             

OXRG02 

Blue 

Line 

Test 

Window 

The OratectXP

®

* Oral Fluid Drug Screen Device is a simple one-step test for the detection of drugs of abuse in oral fluid. 

OratectXP

®

 Procedures -- 

Refer to package insert for detailed instructions and technical information.

 

1.  Remove

 the blue cap by holding 

the sides and pulling gently. This 
will expose the collection pad.  

Make sure there is a blue line 
present in the window area 
indicating an unused device.

 

2.  Open 

mouth and gently 

rub 

the 

collection pad inside mouth against 
cheek in a circular motion several 
(approximately 15-20) times. Make 
sure to 

keep head level.

 

    

 

      

3. 

Gently

 rub

 the collection pad 

against the opposite cheek in 
circular motion several 
(approximately 15-20) times. 

 

  

     

4. 

Gently

 rub

 the collection pad 

on 

top

 of the tongue several 

(approximately 15-20) times.  

Do 

not chew, suck, bite or bend the 
collection pad.

 

5.  Rub

 the collection pad 

underneath

 

the tongue several (approximately 
15-20) times. 

6. 

Place the collection pad underneath 
the tongue for approximately 30 
seconds to collect saliva. Instruct the 
donor to hold the device in place with 
hand.  

Sufficient amount of saliva 

collected is indicated by the flow 
of the blue line.  Repeat steps 2-6 
until the blue line flow.

 

  
 
 
 
 
 
 
 
 

 

7.  Remove

 from mouth as soon as 

the blue line flow at the test 
window.  

Re-cap

 the device. 

 

 

 

 

8. 

Lay the device on a flat surface 
and

 read results in 3 minutes 

after removing the device from 
mouth.  See Package Insert.

 

Interpreting OratectXP

®

 Test Results 

Invalid Result 

When

 no

 colored band appears in the 

CONTROL (C) region

the test is 

invalid 

even if there is a band in the 

test region.  

Repeat

 the test with a new 

device. 
 

 

Negative Result 

Colored bands should be observed: 

 

One in the CONTROL (C) region  

 

One in the specific TEST region 

The color of the test band may be 
slightly darker or lighter than the control 
band. Any visible band that can be seen 
is a 

negative

 result.

 

Presumptive Positive Result 

A colored band at the 

CONTROL (C) 

region

 should be observed.  When 

there is no colored band at the 
specific 

TEST region

, the test is 

presumptive positive for that particular 
drug. 
 

 

Example Interpretation: 

Example Interpretation: 

Example Interpretation: 

 

 

OPI: 

Invalid 

MET: 

Invalid 

THC: 

Invalid 

COC:  Invalid 

 

 

 

OPI: 

Negative 

MET: 

Negative 

THC: 

Negative 

COC:  Negative 

 

 

OPI: 

Negative 

MET: 

Presumptive 

 

Positive 

THC: 

Negative 

COC:  Negative

 

 

 

 

 

 

 

 

 

 

 

 

Reviews: