Test 

Windows 

Test 

Windows

 

The Oratect

® 

 Oral Fluid Drug Screen Device is a simple one-step test for the detection of drugs of abuse in oral fluid. 

Oratect

® 

 Procedures -- 

Refer to package insert for detailed instructions and technical information.

 

1.  Remove

 the blue cap by holding 

the sides and pulling gently. This 
will expose the collection pad.  

Make sure there is a blue line 
present in each window area 
indicating an unused device.

 

2.  Open 

mouth and gently 

rub 

the 

collection pad inside mouth against 
cheek in a circular motion several 
(approximately 15-20) times. Make 
sure to 

keep head level.

 

    

 

      

3. 

Gently

 rub

 the collection pad 

against the opposite cheek in 
circular motion several 
(approximately 15-20) times. 

 

  

 

     

4. 

Gently

 rub

 the collection pad 

on 

top

 of the tongue several 

(approximately 15-20) times.  

Do 

not chew, suck, bite or bend the 
collection pad.

 

5.  Rub

 the collection pad 

underneath

 

the tongue several (approximately 
15-20) times. 

6. 

Place the collection pad underneath 
the tongue for approximately 30 
seconds to collect saliva. Instruct the 
donor to hold the device in place with 
hand.  

Sufficient amount of saliva 

collected is indicated by the flow 
of the blue lines.  Repeat steps 2-6 
until blue lines flow.

 

  

7.  Remove

 from mouth as soon as 

blue lines flow at both of the test 
windows.  

Re-cap

 the device. 

 

 

 

 

8. 

Lay the device on a flat surface 
and

 read results in 5 minutes 

after removing the device from 
mouth.  See Package Insert.

 

Interpreting Oratect

®

 Test Results 

Invalid Result 

When

 no

 colored band appears in the 

CONTROL (C) region

the test is 

invalid 

even if there is a band in the 

test region.  

Repeat

 the test with a new 

device. 
 

 
 
 
 
Example Interpretation: 
 

ME:  Invalid 
TH: 

Invalid 

CO:  Invalid 
AM:  Invalid 
OP:  Invalid 
PC:  Invalid 

 

Negative Result 

For each test, two colored bands should 
be observed: 

 

One in the CONTROL (C) region  

 

One in the specific TEST region 

The color of the test band may be 
slightly darker or lighter than the control 

band. Any visible band 
that can be seen is a 

negative

 result. 

 

Example Interpretation:

 

ME:  Negative 
TH: 

Negative 

CO:  Negative 
AM:  Negative 
OP:  Negative 
PC: 

Negative 

Presumptive Positive Result 

A colored band at the 

CONTROL (C) 

region

 should be observed.  When 

there is no colored band at the specific 

TEST region

, the test is presumptive 

positive for that particular drug. 
 
 

 
 
Example Interpretation: 

 

ME: 

Presumptive 

 

Positive

 

TH: 

Negative 

CO:  Negative

 

AM:  Negative 
OP:  Negative 
PC:  Negative 

 

 

 

 

 

 

 

 

 