BMC AMS42/84-LAN Manual Download Page 3

AMS42/84-LAN 

Page 3 

2

 

Block diagram 

 

3

 

Connections 

All available connectors and operating ele-
ments of the AMS42/84-LAN are accessible 
on the right-hand side of the device front. 
 

3.1

 

Switch 

The AMS42/84-LAN is turned on (press "1") 
and off (press "0") with the switch. The green 
LED indicates the operation of the device. 
 

3.2

 

LEDs 

The two LEDs on the front of the device indicate the following states: 

State 

green LED (ON) 

red LED (LINK) 

flashing 

device on, waiting for connection to software

*

 

transmission of data 

on 

device on, connection to software

*

 connection 

to 

hub 

off 

device off 

no connection to hub 

* NextView

®

4 or LIBAD4; not: iM-Admin 

 

3.3

 

Network 

The PC/notebook and the AMS42/84-LAN communicate via network, which is realized with 
by a Twisted Pair connection (10Mbit). If directly connecting the AMS42/84-LAN to the PC, a 
crosslink cable must be used. Both cables are available as accessories (

ZUKA10T

 or 

ZUKA-

CL10T

). 

 

A firewall software installed on the PC might hinder the operation of the AMS42/84-LAN. In this case, 
please contact your network administrator. 

Summary of Contents for AMS42/84-LAN

Page 1: ...e and be integrated in the AMS Clearly safe The galvanic isolation provided by the 5B mod ules guarantees interference free operation and protects the DAQ system and the PC against high potentials 8 1...

Page 2: ...be prepared first before attaching the device to the rack system with screws 1 2 Software installation and start up procedure a Connect an external power unit to the 3 pole DIN coupling for 9 40V pow...

Page 3: ...K flashing device on waiting for connection to software transmission of data on device on connection to software connection to hub off device off no connection to hub NextView 4 or LIBAD4 not iM Admin...

Page 4: ...sepa rately if possible Measurements can be effected on various potentials The potential deviations however should not ex ceed 60V DC according to VDE Association of German Electrical Engineers 3 6 Sy...

Page 5: ...outputs Serial resistors in the output lines limit the current and protect the hardware from being destroyed 4 1 Connecting an LED Only so called low current LEDs can be used because they already work...

Page 6: ...nment The pin assignment on the right shows the top view of the module back plane see chapter 5 The pin assignment corresponds to the 5B modules of Analog Devices BURR BROWN etc However an additional...

Page 7: ...for C C on Windows MAC OS X FreeBSD Linux SDK LIBAD IG installation programming manual Operating program NV4 measuring software NextView 4 available in the Standalone versions Lite slim version with...

Page 8: ...pant of the network has his own IP address consisting of four groups of numbers To guarantee that a data package arrives at the correct address this IP address must have been given out only once Each...

Page 9: ...ming the AMS42 84 LAN under Max OS X and Unix FreeBSD Linux with C C is done by means of the LIBAD4 programming interface The SDK LIBAD for the respective operating system also Windows is included on...

Page 10: ...ogram NextView 4 different versions Demo Live Lite Professional Analysis Client Server available unlimited number of devices even of different type can be installed except for NextView 4 Lite configur...

Page 11: ...ally equipped with 8 AMS42 LAN or 16 AMS84 LAN or AMS42 LAN with option AMS EXT8 plug in cassettes Each cassette is equipped with the relevant 5B module and suitable input connector The suitable input...

Page 12: ...Frequency accuracy Frequency drift max 100ppm max 50ppm C The total sampling rate is the sum of the sampling rates of the individual used channels e g if 5 channels are scanned with 1kHz the total sa...