Cleaning the Mask at Home 

Notes:

 

 

The mask and headgear can only be cleaned through hand washing or 
placed in the Lumin device. 

 

The elbow and the valve cannot be dissembled for washing. 

 

CAUTIONS

 

 

Do  not  use  solutions  containing  bleach,  chlorine,  alcohol,  aromatics, 
moisturizers, antibacterial agents, or scented oils to clean any part of the 
system or air tubing. These solutions may cause damage and reduce the 
life of the product. 

 

Exposing any part of the system or tubing to direct sunlight or heat may 
cause deterioration.   

 

If any visible deterioration of a component is apparent (cracking, crazing, 
tears etc), the component should be discarded and replaced.   

Daily / After Each Use                                                                                       

 

 

Disassemble  the  mask  components  according  to  the  disassembly 
instructions. 

 

Thoroughly clean the separated mask components (excluding headgear), 
by  gently  rubbing  in  warm  (approx.  30°C)  water  using  mild  liquid 
detergent (e.g., Alconox diluted at 1%) for up to 10 minutes. 

 

Use a soft bristle brush to clean the vent. 

 

Rinse all components well with drinking quality water and allow to air dry 
out of direct sunlight. 

 

When all components are dry, reassemble according to the reassembly 
instructions. 

Weekly                                                                                                                 

 

Handwash  the  headgear  and  all  components  in  warm  (approx.  30°C)  and 
mild liquid detergent (e.g., Alconox diluted at 1%) for up to 10 minutes. Rinse 
the components well with drinking quality water and allow them to air dry out 
of direct sunlight before assembling. 

Reassembling the Mask 

 

 

 

Install the forehead pad 
onto 

the 

upper-half 

installation  groove  of 
the mask frame. 

Install the Cushion onto 
the Frame and confirm 
that the Cushion is 
reliable. 

Install the upper-half 
headgear onto the 
upper-half installation hole 
of the mask frame. 

Technical Specifications 

Problem 

Possible Reason 

Possible Solution 

Mask won’t seal 
properly or is 
uncomfortable. 

Mask may have been 
fitted incorrectly. 

Carefully follow instructions in 
“Fitting the Mask” section. Make 
sure the headgear is not 
over-tightened.   

Mask size is wrong. 

Consult your clinician. 

Mask leaks around 
the face. 

The cushion is 
misplaced on the 
cushion frame. 

Check insertion of the cushion and 
reinsert correctly according to the 
instructions in 

“Reassembling the 

Mask

” section.   

Mask is too noisy. 

Mask size is wrong. 

Consult your clinician. 

Vents are blocked or 
partially blocked. 

Clean the vents according to the 
instructions in “Cleaning the Mask 
at 

Home” section.   

Technical Specifications 

Pressure

–Flow Curve 

 

 

 

Pressure

cmH

2

O

 

12 

22 

31 

40 

Flow Rate

L/min

 

19 

41 

57 

69 

79 

 

Dead Space 
Information

 

Dead space is the empty volume of the mask up to the swivel.   
The dead space of the mask varies according to cushion sizes 
but is less than 225 mL. 

Therapy Pressure 

3 to 40 cmH

2

Resistance   

Drop in Pressure measured (average for 3 sizes) 
at 50 L/min: 0.15 cmH

2

at 100 L/min: 0.5 cmH

2

Inspiratory and 
Expiratory Resistance 

The  inspiratory  resistance  of  the  mask  (in  combination  with 
the Non-Rebreathing Valve) is 1.8 cmH

2

O at 50 L/min. 

The expiratory resistance of the mask (in combination with the 
Non-Rebreathing Valve) is 2.0

 

cmH

2

O at 50 L/min. 

The Non-Rebreathing 
Valve 
open-to-atmosphere 
pressure 

  2.8 cmH

2

The Non-Rebreathing 
Valve 
closed-to-atmosphere 
pressure 

  2.8 cmH

2

Sound 

DECLARED DUAL-NUMBER NOISE EMISSION VALUES in 
accordance  with  ISO  4871.  The  A-weighted  sound  power 
level  of  the  mask  is  28  dBA,  with  uncertainty  3  dBA.  The 
A-weighted sound pressure level of the mask at a distance of 
1 m is 20 dBA, with uncertainty 3 dBA. 

Environmental 
Conditions 

Operating temperature: +5°C to +40°C (41°F to 104°F) 
Operating  humidity:  10% 

  93%  relative  humidity 

non-condensing 
Storage and transport: -20°C to +55°C (-4°F to 131°F) 
Storage  and  transport  humidity:  10% 

  93%  relative 

humidity non-condensing 

Storage 

Ensure  that  the  mask  is  thoroughly  clean  and  dry  before  storing  it  for  any 
length of

 

time. Store the mask in a dry place out of direct sunlight. 

Disposal 

The mask does not contain any hazardous substances and may be disposed 
of with your normal household refuse. 

Symbols 

System and Packaging 

 

Caution, consult accompanying documents. 
Indicates a Warning or Caution and alerts you to a possible injury or 
explains  special  measures  for  the  safe  and  effective  use  of  the 
device. 

 

Lot number 

 

Catalog number 

 

Temperature limitation 

 

Humidity limitation 

 

Manufacturer 

 

Authorized Representative in the European Community 

Limited Warranty   

It  is  warranted  that  the  Siesta  Full  Face  Mask,  including  Frame  assembly, 
Cushion, Forehead pad, Headgear shall be free from defects in material and 
workmanship for a period of ninety (90) days from the date of purchase by the 
initial consumer. 
To  exercise  the  rights  under  this  warranty,  contact  the  local  authorized 
dealers. 
The expected service life of Siesta Full Face Mask is one year. 

For  any  questions  or  comments  concerning  this  product  or  the  warranty, 
please contact: 

3B Medical 
203 Avenue A NW, Suite 300 
Winter Haven, FL 33881 
Email: [email protected] 

www.3Bproducts.com

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
391124                                                                                               