Clinician Set-Up Menu
Review Prior Night Summary
Display
Screen
User
Button (+)
Ramp Button
Pressure Start/Stop
Button
User
Button (–)
Heated
Humidifier
Button
To enter the Clinician’s Set-Up Menu:
Remove the power cord from the rear of the RESmart
®
unit. Press and hold both the Start and Ramp buttons
simultaneously, while reconnecting the power cord.
To Select an Option:
Press the Ramp button once. The cursor will blink, and values may
be changed using the user b/-.
Quickly press and release user buttons.
To Confirm a Setting:
Press the Ramp button again and the cursor will stop blinking
To Advance to the next Option:
Press the User + or - button to move through the Clinician Menu
To SAVE Menu Selections:
Press the Ramp button when the display screen prompts “SAVE”
NOTE: Changes to any options will be final only when you have
confirmed them by pressing the RAMP button when you are at the
SAVE selection on the display screen.
Reviewing Sleep Data
The RESmart
®
allows review of the prior night’s sleep data. To access your data, make sure that the
blower is not running by pressing the stop/start button.
Press the Ramp Start button , and scroll through the display using the +/- buttons. The data
displayed consists of completion time (Compl), P95 (95th percentile of the airway pressure over
time), AHI (apnea-hypopnea index), and SNI (snore index).
For a discussion on interpreting this data, contact your physician or home care provider.
TREAT P: (CPAP mode) treatment pressure may be
adjusted from 4 cmH
2
O to 20 cmH
2
O in 0.5 cmH
2
O
increments.
MIN APAP: The setting is the minimum pressure of
the therapy pressure range in Auto Mode.
MAX APAP: The setting is the maximum pressure
of the therapy pressure range in Auto Mode.
RAMP P: The starting pressure after blower is
turned on when RAMP Time is set at anything other
than 0 Minutes. The setting range is 4 cmH
2
O to
treatment pressure (in CPAP Mode) or MIN APAP (in
AUTO Mode), and the adjustment step is 0.5 cmH
2
0.
USE TIME: Total hours this device has been used.
This number is only for reference.
MASK TEST: A calibration feature which can
be used to ensure the patient is wearing the mask
correctly and to detect leaks.
SAVE AND EXIT: Press the Ramp button to
confirm and SAVE all settings.
3B Medical, Inc.
799 Overlook Drive • Winter Haven, FL 33884
(863) 226-6285 • [email protected]
www.3Bproducts.com