Clinician Set-Up Menu

Review Prior Night Summary

Display

Screen

User

Button (+)

Ramp Button

Pressure Start/Stop

Button

User

Button (–)

Heated

Humidifier

Button

To enter the Clinician’s Set-Up Menu:

    

Remove the power cord from the rear of the RESmart

®

 

unit.  Press and hold both the      Start and       Ramp buttons 
simultaneously, while reconnecting the power cord.

To Select an Option:

    Press the Ramp button once.  The cursor will blink, and values may 
be changed using the user b/-.
    

Quickly press and release user buttons.

To Confirm a Setting: 

    Press the Ramp button again and the cursor will stop blinking

To Advance to the next Option:  

    Press the User + or - button to move through the Clinician Menu 

To SAVE Menu Selections:

  Press the Ramp button when the display screen prompts “SAVE”  

NOTE:  Changes to any options will be final only when you have 
confirmed them by pressing the RAMP button when you are at the 
SAVE selection on the display screen. 

Reviewing Sleep Data

The RESmart

®

 allows review of the prior night’s sleep data. To access your data, make sure that the 

blower is not running by pressing the stop/start button.      

Press the Ramp Start button       , and scroll through the display using the +/- buttons. The data 
displayed consists of completion time (Compl), P95 (95th percentile of the airway pressure over 
time), AHI (apnea-hypopnea index), and SNI (snore index).

For a discussion on interpreting this data, contact your physician or home care provider.

TREAT P:  (CPAP mode) treatment pressure may be 
adjusted from 4 cmH

2

O to 20 cmH

2

O in 0.5 cmH

2

increments.
MIN APAP:  The setting is the minimum pressure of 
the therapy pressure range in Auto Mode.
MAX APAP:  The setting is the maximum pressure 
of the therapy pressure range in Auto Mode.
RAMP P:  The starting pressure after blower is 
turned on when RAMP Time is set at anything other 
than 0 Minutes. The setting range is 4 cmH

2

O to 

treatment pressure (in CPAP Mode) or MIN APAP (in 
AUTO Mode), and the adjustment step is 0.5 cmH

2

0.

USE TIME:  Total hours this device has been used. 
This number is only for reference.
MASK TEST:  A calibration feature which can 
be used to ensure the patient is wearing the mask 
correctly and to detect leaks. 
SAVE AND EXIT:  Press the Ramp       button to 
confirm and SAVE all settings. 