To Enter the Patient Option Menu:
Press the Ramp button for 3 seconds
To Select an Option:
Press the Ramp button once. The cursor will blink, and values may be
changed using the User b/-.
Quickly press and release user buttons.
To Confirm a Setting:
Press the Ramp button again and the cursor will stop blinking
To Advance to the next Option:
Press the User + or - button to move through the Patient Menu
To SAVE Menu Selections:
Press the Ramp button when the display screen prompts “SAVE”
NOTE: Changes to any options will be final only when you have
confirmed them by pressing the RAMP button when you are at the
SAVE screen.
Quick Reference Sheet
RESmart
®
CPAP & Auto-CPAP
3135_10.2016
Patient Menu Set-Up
Setting Options:
HEATER: When the optional heated humidifier is
used, this setting can change the temperature and
humidity of airflow between level 0 to 5, ranging
from lowest to highest setting.
ALTITUDE: The RESmart CPAP and Auto-CPAP
devices are auto adjusting. The device may be
optimized at higher altitudes by setting the
altitude as follows:
0 - less than 2,460 feet
1 - 2,460 to 4,921 feet
2 - 4,921 to 8,202 feet
AUTO ON: Auto On is a setting to enable an
inspiration trigger on the device. The device will
start airflow immediately after a breath is taken
wearing a mask. Options are ON or OFF. Default
setting is ON.
AUTO OFF: When the mask is removed, the
device will stop airflow automatically. Options are
ON or OFF. Default setting is OFF
LIGHT: The LCD light may be set to remain on
always, remain dimmed, or automatically go
bright when a button is pressed and remain bright
for a few seconds. Options are ON, OFF, and
AUTO. Default setting is AUTO.
ALERT: The audible alert for leak and power off
may be enabled or disabled. Default setting is OFF.
RESLEX
™
: RESlex
™
refers to BMC’s exhalation
relief feature, standard on all models. Exhalation
relief allows a pressure drop during exhalation
for patient comfort. Available options range from
0 to 3, with 3 affording the greatest exhalation
pressure relief. Default setting is 0.
RAMP TIME: This setting is the ramp timer
from starting pressure to prescription pressure.
Setting range is 0-60 mins with an adjustment step
of 5 min.
DELAY OFF: When the integrated heated
humidifier is used, this setting allows the air flow
to continue for an additional 15 minutes at very
low pressure after the unit is turned off. This
option is intended to dry off moisture, extend the
operating life of the device by reducing humidity,
and dry out tubing. Options are ON or OFF.
Default setting is OFF.
SETTING THE DATE: The current date is
displayed and can be modified if necessary.
SETTING THE TIME: The time can also be
set by this option and is in 24 hours format. It
is important to set the correct time as this will
impact accuracy of sleep reports.
USE DAYS: Use days refers to the number of days
compliant (i.e. days where device use > 4 hours).
This is for quick reference only.
SAVE: Press Ramp button to confirm the
saving of all selections.
Display
Screen
User
Button (+)
Ramp Button
Pressure Start/Stop
Button
User
Button (–)
Heated
Humidifier
Button