To Enter the Patient Option Menu:

   Press the Ramp      button for 3 seconds

To Select an Option:

    Press the Ramp button once.  The cursor will blink, and values may be 
changed using the User b/-.
    

Quickly press and release user buttons.

To Confirm a Setting: 

    Press the Ramp button again and the cursor will stop blinking

To Advance to the next Option:  

    Press the User + or - button to move through the Patient Menu 

To SAVE Menu Selections:

    Press the Ramp button when the display screen prompts “SAVE”  

NOTE:  Changes to any options will be final only when you have 
confirmed them by pressing the RAMP button when you are at the 
SAVE screen. 

Quick Reference Sheet

RESmart

®

 CPAP & Auto-CPAP

3135_10.2016

Patient Menu Set-Up

Setting Options:

HEATER: When the optional heated humidifier is 
used, this setting can change the temperature and 
humidity of airflow between level 0 to 5, ranging 
from lowest to highest setting.
ALTITUDE: The RESmart CPAP and Auto-CPAP 
devices are auto adjusting. The device may be 
optimized at higher altitudes by setting the 
altitude as follows:

 

 

0 -  less than 2,460 feet 

 

 

1 - 2,460 to 4,921 feet

 

 

2 - 4,921 to 8,202 feet 

AUTO ON:  Auto On is a setting to enable an 
inspiration trigger on the device. The device will 
start airflow immediately after a breath is taken 
wearing a mask. Options are ON or OFF. Default 
setting is ON.
AUTO OFF: When the mask is removed, the 
device will stop airflow automatically. Options are 
ON or OFF.  Default setting is OFF
LIGHT: The LCD light may be set to remain on 
always, remain dimmed, or automatically go 
bright when a button is pressed and remain bright 
for a few seconds. Options are ON, OFF, and 
AUTO. Default setting is AUTO.
ALERT: The audible alert for leak and power off 
may be enabled or disabled. Default setting is OFF.

RESLEX

: RESlex

 refers to BMC’s exhalation 

relief feature, standard on all models. Exhalation 
relief allows a pressure drop during exhalation 
for patient comfort. Available options range from 
0 to 3, with 3 affording the greatest exhalation 
pressure relief. Default setting is 0.
RAMP TIME:  This setting is the ramp timer 
from starting pressure to prescription pressure. 
Setting range is 0-60 mins with an adjustment step 
of 5 min.
DELAY OFF:  When the integrated heated 
humidifier is used, this setting allows the air flow 
to continue for an additional 15 minutes at very 
low pressure after the unit is turned off. This 
option is intended to dry off moisture, extend the 
operating life of the device by reducing humidity, 
and dry out tubing. Options are ON or OFF. 
Default setting is OFF.
SETTING THE DATE: The current date is 
displayed and can be modified if necessary. 
SETTING THE TIME: The time can also be 
set by this option and is in 24 hours format. It 
is important to set the correct time as this will 
impact accuracy of sleep reports.
USE DAYS: Use days refers to the number of days 
compliant (i.e. days where device use > 4 hours). 
This is for quick reference only.
SAVE:  Press Ramp       button to confirm the 
saving of all selections. 

Display

Screen

User

Button (+)

Ramp Button

Pressure Start/Stop

Button

User

Button (–)

Heated

Humidifier

Button