4.1.3
PRECAUTION ASSOCIATED WITH EQUIPMENT CARE AND MAINTENANCE
The device must not be altered in any way without the manufacturer’s prior consent: the alteration
could lead to unpredictable operation, potentially injuring the patient and caregivers.
Torn upholstery no longer presents an effective antibacterial barrier and must hence be replaced
as soon as possible.
The device must be regularly inspected, at least once per year. See the maintenance check-list.
To maintain the device in good condition and to guarantee its safe and durable operation, the
prescribed maintenance must be performed.
In the event of observed anomalies or damage that could jeopardize device operation and
safety, the equipment must be immediately withdrawn.
The device should only be used according to the instructions in this user manual.
The device must not be exposed to, or even less be in contact with, sources of heat, combustion,
and flammable agents.
Work on electrical parts should only be performed by qualified and authorized staff.
* BMB MEDICAL will accept no liability in the event of incidents or accidents resulting from
improper use or alteration of the product by the buyer or user.
4.1.4
MC-RELATED INFORMATION
The products in the I-MOVE range are suitable for use in all healthcare establishments, with the
exception of facilities in the vicinity of HF ELECTROSURGICAL APPLIANCES, or near an
electromagnetically shielded room for an EM magnetic resonance imaging system, in which the
intensity of EM INTERFERENCES is high.
The products in the I-MOVE range are not suitable for use in domestic homes or establishments
or on any premises connected to the public low-voltage network supplying domestic buildings.
EM interference could cause the product to make unexpected movements.
This appliance should not be used next to, or stacked on other appliances as this could lead
to incorrect operation. If this mode of operation is necessary, this and the other appliances
should be observed to ensure their correct operation.
The use of accessories, transducers or cables other than those specified or supplied by the
manufacturer can lead to an increase in electromagnetic emissions, or a decrease in this
appliance’s immunity, causing inappropriate operation.
Do not use portable RF communication devices (including peripherals such as antenna
cables and external antennas) less than 30 cm (12 inches) from any part of products in the
I-MOVE range, including the cables specified by the manufacturer. Otherwise, the
performance of these devices could be impaired.