Tell the patient and any accompanying persons not to sit on the armrests.
•
The equipotential connector must not, under any circumstances, be used as a protective
earth connector
•
Failure to perform these cleaning operations can lead to a risk of infection caused by
the presence of secretions and/or residues on the chair.
•
Before cleaning the chair, unplug it from the mains and ensure that all internal connections
are correctly established.
•
High-pressure and steam cleaning are prohibited.
•
Any use that does not fall within the operating context of a hospital chair.
•
To prevent failure, electromedical equipment requires special
•
Never use the chair in the presence of an anesthesia mixture flammable with air, with
oxygen, or with nitrous oxide as this could cause a fire and/or explosion.
•
Use by any non-caregiver who has not received the authorization and explanations from
a qualified individual.
All manipulations relating to the aforementioned precautions for use are described in
paragraph II - Instructions for use
.
3.4.2
CONTRAINDICATIONS
There are no noteworthy contraindications for this device.