23 

www.BlueBoxOptics.com V2.4 
 

 

niji  

 

Figure 13: Filter Positioning 

Note: Every effort should be made to avoid contaminating the surface of the optical 
filters. Should a filter become contaminated with dust or grease, it can be cleaned.  
Full instructions on cleaning are available on our website, 

www.blueboxoptics.com

 

To replace a filter on any channel, with the exception of the Green/Amber channel, 
position the filter paddle rough side down on a smooth and clean surface. Remove 
the screw and set to one side. Gently release the filter by pushing the filter from the 
smooth to the rough side of the paddle. Do not make contact with the surface of the 
filter itself, but instead apply force to the outer black mounting ring. Before inserting a 
filter, position the paddle rough side up on a smooth and clean surface. Each filter 
has a small arrow printed onto its edge, place the filter into the paddle with the arrow 
pointing towards the rough side. Avoid making contact with the coated surface of 
the filter and instead apply force to the outer black mounting ring. Replace the screw 
and tighten using the supplied 1.5mm hex key to secure the filter in the paddle (Figure 
14). 

Summary of Contents for niji

Page 1: ...User Manual niji...

Page 2: ...Driver Setup USB control only 7 3 4 Software Setup 7 3 5 Hardware Placement and Ventilation 7 3 6 Hardware Set Up 8 3 7 Connect to the niji 9 3 8 Software Control 10 3 9 Constant Optical Power Mode 1...

Page 3: ...and String 28 8 3 Channel Level Set Example 28 8 4 Channel On Off 29 8 5 Channel Active Set Example 29 8 6 Channel Identification 29 8 7 TTL Command String 29 8 8 TTL Set Example 29 9 Technical Specif...

Page 4: ...17 Figure 9 Preset Options 18 Figure 10 Preset Loaded 19 Figure 11 Internal Temperature Display 20 Figure 12 niji Filter Paddle 22 Figure 13 Filter Positioning 23 Figure 14 A Filter and Filter Paddle...

Page 5: ...cation of this equipment will invalidate the warranty and may render the unit inoperable Switch off the unit when not in use Do not allow liquids to enter the housing Do not cover air vents Do not sta...

Page 6: ...where these exist Registration WEE BU4061XA 1 4 Static Sensitivity Appropriate precautions should be taken to prevent damage from static electricity discharge 1 5 Technical Support For technical supp...

Page 7: ...ision Light output can be coupled into a liquid light guide LLG The niji system also comprises a TTL port whereby a TTL signal can trigger light output switching on off The niji multi wavelength light...

Page 8: ...plication allows the user to control each individual channel software is provided with each unit and written exclusively for Bluebox Optics 3 3 Driver Setup USB control only The drivers for this unit...

Page 9: ...al temperature probes are installed to monitor the equipment temperature and this data is displayed via the software interface 3 6 Hardware Set Up Connect the liquid light guide LLG to the device via...

Page 10: ...several built in safety conditions which must be satisfied before any light is emitted When the power indicator at the front of the unit flashes red the following actions should be taken Ensure that...

Page 11: ...s they are in Figure 4 When a single or multiple channels are switched on the status will change to LIGHTS ON This is an integral safety feature by design See Table 1 for all available light source co...

Page 12: ...vely Figure 4 Lights OFF All Channels OFF Figure 5 Lights ON All Channels On Seven channels are available Channel 5 delivers highly broadband light and as standard is configured in the broadest band a...

Page 13: ...m to low optical power applciations This can be enabled in the Configuration window In Constant Optical Power mode the niji can be controlled using the sliders etc as it would normally be used but the...

Page 14: ...13 www BlueBoxOptics com V2 4 niji...

Page 15: ...programmable with many microscope systems Custom configuration including intial power on settings can be achieved using the XML configuration file If you wish to use the niji in compatible mode or wi...

Page 16: ...off when 5V is applied The level of emission of each channel is set either through the niji software or through serial commands Those channels that are not connected to the TTL interface will remain...

Page 17: ...the TTL triggered channels will be controlled by the TTL interface In this mode only those channels which are tied to 5V are in the off state Active High Active Low 100k Pull Up 100k Pull Down 5V GND...

Page 18: ...17 www BlueBoxOptics com V2 4 niji Figure 8 TTL Mode Active...

Page 19: ...be saved and recalled at any time however an unlimited number of preset files can be saved if needed see Figure 91 Figure 9 Preset Options To save a preset select your chosen preset identification na...

Page 20: ...reset Loaded 3 13 Software Dual Operation Mode The niji can receive and send commands to multiple sources via the separate hardware connection ports USB and RS232 For example with the niji software ru...

Page 21: ...ternal Temperature The internal temperature of the unit is displayed in the information tab Case Temperature 2 monitors ambient case temperature Case Temperature 1 monitors the temperature at board le...

Page 22: ...level where 100 represents the maximum available intensity and 1 represents the minimum emission of the system 4 3 Light Guide Adapter and Interlock To install the light guide push the 3mm light guid...

Page 23: ...dle shaped material which is safe to handle Figure 12 Figure 12 niji Filter Paddle One side of the paddle is rough while the other is smooth The smooth surface should face towards the module with the...

Page 24: ...w and set to one side Gently release the filter by pushing the filter from the smooth to the rough side of the paddle Do not make contact with the surface of the filter itself but instead apply force...

Page 25: ...gure 15 Place the paddle on a smooth and clean surface remove the screw and set to one side Gently release the filter by pushing through from the smooth side of the paddle Before inserting a filter po...

Page 26: ...replace the lid as soon as possible Each module is labelled with its channel number and name indicating its location in the niji This labelled area also acts as a grip for the removal and insertion o...

Page 27: ...made via a card edge connection this will need to be gently eased out Do not allow thumb screws to roll into the unit from the top plate Set them to the side once removed Store any unused modules in...

Page 28: ...cleaning procedure should the surface of the filters be contaminated through handling exposure to dust etc Full instructions can be found on our website www blueboxoptics com 8 Serial Command Control...

Page 29: ...port should be configured as follows Port select COM Baud rate 115200 Data bits 8 Stop Bit 1 Parity None Flow control RTS CTS Forward None Transmitted CR and LF Figure 19 Port Setup 8 2 Channel Level...

Page 30: ...avelength 1 Ultra Violet 2 Royal Blue 3 Blue 4 Cyan 5 Green Amber 6 Red 7 Far Red 8 7 TTL Command String The TTL command string is split into four parts These are the command identifier function ON OF...

Page 31: ...Mercury Free Yes Operating Conditions 10 to 40 C Storage Temperature 10 to 60 C Humidity 40 to 70 RH Warranty 12 Months 3 year Optional 9 1 External Power Supply Manufacturer Bluebox Optics Ltd Model...