9
7
Static Alignment
2.5
Fully extend limb to access the adjuster.
Rotate the SPRING STACK ADJUSTER to
achieve the required stability for the individual
amputee.
Note… Audible clicks should be heard upon
adjustment, and the adjuster is self locking.
Tighten the SPRING STACK to reduce stability.
Release the SPRING STACK to increase stability
- See above.
If required, the adjuster may be reset to
the factory setting. This can be achieved by
reducing stability by six audible clicks from the
maximum stability setting.
The instructions in this section are for practitioner use only.
Decrease
Stability
Increase
Stability
Walking Trials
Underside View of Chassis
Build Line
Trochanter
Knee Pivot
Ankle Centre
Spring Stack Adjuster