54
Chapter 4
December 2015
Flex Focus 1202 User Guide
(BB1756-Q)
The display on the back of the battery charger station shows % of battery capacity
(see Fig 4-2). A display of
::::
indicates there are no batteries in the battery charger
station.
Battery
indicators
When the door to the battery charger station is open, you can see LED indicator lights
showing the status of each individual battery (see Fig 4-4).
Recharging a Battery in the Battery Charger Station
To insert or remove batteries:
1
Press the latch on the back side of the battery charger station to open the cover.
2
Slide drained (or new) batteries into the empty slots.
3
When a battery is fully charged, pull its tab and slide it completely out of the
frame.
NOTE:
See the battery disposal warning (BS-4) in “Battery Support System” on
NOTE:
See the warning about using only authorized UA1225 batteries (BS-2) in
“Battery Support System” on page 29.
Cleaning and Disinfection
The cleaning and disinfection instructions for the system in
Care, Cleaning & Safety
apply to the battery charger station.
LED Indicator Light
Battery Status
Green light
Charged
Red light
Battery level is below 33%
Off
No battery in slot or battery is flat
Summary of Contents for Flex Focus 1202
Page 5: ...5 Chapter 6 Getting Started 59 Index 61 ...
Page 6: ...6 ...
Page 10: ...10 Chapter 1 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 44: ...44 Chapter 2 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 46: ...46 Chapter 3 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 58: ...58 Chapter 5 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 60: ...60 Chapter 6 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 64: ...64 ...
Page 65: ......
Page 66: ......