The sterile-packed needle guide must be stored at a temperature range from +5ºC
(+41ºF) to +25ºC (+77ºF) and a storage humidity of 0% to 80%.
.
Figure 3. Single-use needle guide UA1339.
To take a biopsy:
1. Enclose the transducer in a transducer cover.
2. Click the needle guide into place.
3. Insert a needle into the needle guide.
Ensure the needle guide is securely in place and the tip of the needle guide lies
close to the front of the transducer.
The minimum needle length is 25 cm for a biopsy gun with needle guides
UA1339 and UA1349.
Performing Puncture and Biopsy
WARNING
Sterile-w2
Sterile-packed components must be stored in a safe environment and kept out of direct
sunlight. Large temperature changes during storage may cause condensation and violate
the integrity of the packaging.
WARNING
D-w1
For disposal of contaminated items such as transducer covers or needle guides or other
disposable items, follow disposal control policies established for your office, department, or
hospital.
WARNING
P-w1
Before you start imaging, verify that the type number or name of the transducer and the
type number or description of the needle guide you are using match the number displayed
on the monitor. Also make sure that the needle guide is positioned correctly. If the numbers
do not match, or if the needle guide position is not correct, the puncture line on the monitor
may not correspond to the true puncture path in the tissue. In case of any inconsistency,
stop imaging, turn off the system, and contact your BK service representative.