26
Test Frequency: 1 KHz
Range
Maximum
Display
Accuracy
Spec. Note
Capacitance
DF
1mF
1.999mF *5
3.0%+5
(DF<0.1)
10%+100/Cx+5
(DF<0.1)
After short cal.
200μF
199.99μF
1.0%+5
(DF<0.1)
2.0%+100/Cx+5
(DF<0.1)
After short cal.
20μF
19.999μF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
2000nF
1999.9nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
200nF
199.99nF
0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
20nF
19.999nF
0.7%+5
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
After open cal.
2000pF
1999.9pF
1.0%+5
(DF<0.1)
2.0%+100/Cx+5
(DF<0.1)
After open cal.
Notes:
1.
Q Value is the reciprocal of DF.
2.
This specification is based on the measurement performed at the test socket.
3.
DUT & Test leads should be properly shielded to GUARD if necessary.
4.
Cx=Counts of displayed C value, e.g. C=88.88
μF then Cx=8888.
5.
This reading can be extended to 1999 MAX display with accuracy not
specified.
6.
This reading can be extended to 19999 MAX display with accuracy not
specified.
NOTE:
Specifications and information are subject to change without
notice. Please visit
www.bkprecision.com
for the most current product
information.