Figure 2 - -30 VDC mean(-) 

10.

 

If you adjusted according to the above, you should be able to calculate the mean(+) – mean(-) 
value to be 60 V (30 – (-30) = 60). 

11.

 

Once finished, turn the CH1 volt/div knob to set to 5.00 V scale.  Keep timebase at 2.50 us.  
Repeat steps 6-10, except with 15 V signal from the calibrator, adjust mean(+) to 15.00V and 
mean (-) to –15.00 V to calculate the difference equal to 30 V. 

12.

 

Again, when finished repeat step 11 for each and all other vertical scale ranges accordingly listed 
below with their corresponding settings: 

Volt/div Scale 

DC Signal 

(Set on 

Calibrator)(Positive 

DC/Negative DC)

 

Mean(+) 

Mean(-) 

Mean 

Difference 

(Mean(+) – (Mean(-))

 

2.00 V 

+6VDC/-6VDC 

6.00 V 

-6.00 V 

12 V 

1.00 V 

+3VDC/-3VDC 

3.00 V 

-3.00 V 

6 V 

500 mV 

+1.5VDC/-1.5VDC 

1.50 V 

-1.50 V 

3 V 

200 mV 

+0.6VDC/-0.6VDC 

600.0 mV 

-600.0 mV 

1.2 V 

100 mV 

+0.3VDC/-0.3VDC 

300.0 mV 

-300.0 mV 

600 mV 

50.0 mV 

+0.15VDC/-0.15VDC 

150.0 mV 

-150.0 mV 

300 mV 

20.0 mV 

+0.06VDC/-0.06VDC 

60.00 mV 

-60.00 mV 

120 mV 

10.0 mV 

+0.03VDC/-0.03VDC 

30.00 mV 

-30.00 mV 

60 mV 

5.00 mV 

+15mVDC/-15mVDC 

15.00 mV 

-15.00 mV 

30 mV 

2.00 mV 

+6mVDC/-6mVDC 

6.00 mV 

-6.00 mV 

12 mV 

 

13.

 

Once adjustments have been made for ALL of the vertical scale ranges, press the “SAVE” soft 
button from the menu and all the adjustments will be stored into the instrument. 

14.

 

Now, repeat steps 1-13 for CH2 for re-adjustment. 

15.

 

After both channels are re-adjusted, the instrument needs to do a self calibration prior to use.  To 
do this, press UTILITY and go to Page 2/4 and select “Do Self Cal”.  With nothing connected at 
all the inputs (This is very important as it will affect calibration and all re-adjustments made prior 
in above steps), press SINGLE button to begin self calibration.  This process can take up to a few 
minutes. 

 

Note:   

-

 

Be sure to save before exiting calibration menu. 

-

 

Any key press not mentioned in the above instructions will exit the calibration menu without 
saving any changes.  Keep this in mind while making adjustments for each scale. 

 

Summary of Contents for 2530B

Page 1: ...Model 2530B 2532B Calibration and Adjustment Procedure V111011...

Page 2: ...Revision History Date Edition Revise Reason Revise Content Revised by 11 04 10 V1 0 None None Jason 11 10 11 V111011 Complete re edit All Jeremy...

Page 3: ...complete users will need to lock access to the calibration menu Instructions to do this will be shown at the end of this document After unlocking access to the calibration menu the user must make note...

Page 4: ...ill call this mean For positive DC 7 Next change the DC signal on the calibrator to 30 V 8 Make note of the measured Mean value again We will call this mean For negative DC 9 Now calculate and see if...

Page 5: ...0 mV 150 0 mV 300 mV 20 0 mV 0 06VDC 0 06VDC 60 00 mV 60 00 mV 120 mV 10 0 mV 0 03VDC 0 03VDC 30 00 mV 30 00 mV 60 mV 5 00 mV 15mVDC 15mVDC 15 00 mV 15 00 mV 30 mV 2 00 mV 6mVDC 6mVDC 6 00 mV 6 00 mV...

Page 6: ...ce 2 Press MEASURE button two times 3 Press ACQUIRE button three times 6 If you did it correctly you will enter the calibration menu shown below If you did not redo the above steps again 7 Select Trig...

Page 7: ...14 Now select center from the soft menu and use the universal knob to increase or decrease the value until you catch the critical point between where the signal is triggered and not triggered You wil...

Page 8: ...oints signal is no longer triggered and will move across the screen o 3 12V 7 04V 2 5 08 V This will be the current trigger center level o Turn the LEVEL knob until trigger level is set to 5 08 V leve...

Page 9: ...and observe the trigger range If the trigger range exceeds the waveform display range meaning that the signal is still triggered when trigger level is beyond the waveform display level then the delta...

Page 10: ...ument needs to do a self calibration prior to use To do this press UTILITY and go to Page 2 4 and select Do Self Cal With nothing connected at all the inputs This is very important as it will affect c...

Page 11: ...TILITY button 2 Select Do Self Cal on Page 2 4 of the menu and you will see the black screen that prompts for initiating self calibration 3 At this time do the following in order 1 Press DEFAULT SETUP...