Vertical Controls
24
Note:
Push the
Vertical Variable Knob
to quickly switch between
Coarse
and
Fine
adjustment.
3.2.4 Probe
Sets the probe attenuation factor to match the type of probe being used.
1. Press the CH button of the channel to be configured.
2. Use the softkeys to select
Probe
.
3. Use the softkeys to select
Probe
once more.
4. Use the
Universal Knob
to select the probe attenuation.
Table
shows the probe attenuation factors.
Setting
Description
0.1X
.01 : 1
0.2X
.02 : 1
0.5X
.05 : 1
1X
1 : 1
2X
2 : 1
5X
5 : 1
10X
10 : 1
...
...
10000X
10000 : 1
Table 3.1
Attenuation Factor
To customize the probe attenuation factor:
Press the
Probe
softkey, select
Custom
, and then press the Custom softkey.
Use the
Universal Knob
to set the desired probe attenuation ratio.
The range is [1E-6,1E6].
3.2.5 Unit
Selects the amplitude display unit for the selected channel.
The available units are
V
and
A
.
1. Press the CH button of the channel to be configured.
2. Use the softkeys to navigate to page 2/2.
3. Use the softkeys to select
Unit
and alternate between
V
and
A
.
The default unit is
V
.