34
APPLICATIONS_______________________________________________________________________________________
Two Differing Negative Voltages With a Positive
Common (Refer To Fig. 14)
Using both supplies, two differing negative voltages with a
common positive are obtained as follows:
1.
Select the
INDEP
endent operating mode and set the A/B
Metering switch to monitor the “A” supply.
2.
Set the desired voltage and maximum current for the “A”
supply using the “A” VOLTAGE and CURRENT
controls.
3.
Set the A/B Metering switch to monitor the “B” supply.
4.
Set the desired voltage and maximum current for the “B”
supply using the “B” VOLTAGE and CURRENT
controls.
5.
Connect the ground wires between each supplies’
positive terminal and GND (green) terminal.
6.
Turn off the power supply and the equipment to be
powered during hook-up.
7.
Connect the negative polarity inputs of the circuit to be
powered to the negative (black) terminals of the supplies.
8.
Connect the common positive input of the circuit to be
powered to either supplies’ positive (red) or the GND
(green) terminal.
Fig. 14. Typical Hook-Up Using Two Different
Negative Voltages and a Common Positive.
Summary of Contents for 1651A
Page 54: ...50 Figura 1 Controles e indicadores del panel frontal...
Page 57: ...53 Figura 2 Controles del panel trasero...
Page 60: ...56 Figura 3 Posibilidades de conexi n a tierra en operaci n independiente 1 de 2...
Page 61: ...57 Figura 3 Posibilidades de conexi n a tierra en operaci n independiente 2 de 2...
Page 65: ...61 Fig 8 Posibilidades de conexi n a tierra en la operaci n de rastreo en serie 0 48V 1 de 3...
Page 66: ...62 Fig 8 Posibilidades de conexi n a tierra en la operaci n de rastreo en serie 0 48V 2 de 3...
Page 69: ...65 Fig 9 Posibilidades de conexi n a tierra en operaci n de rastreo en paralelo 1 de 2...
Page 70: ...66 Fig 9 Posibilidades de conexi n a tierra en operaci n de rastreo en paralelo 2 de 2...
Page 72: ...68 Fig 10 Posibilidades de conexi n a tierra para la fuente fija de 5V 1 de 2...
Page 73: ...69 Fig 10 Posibilidades de conexi n a tierra para la fuente fija de 5V 2 de 2...