2052 User Guide
(BB1804-C)
Figure 6. UA0048 water standoff system components.
NOTE:
See the Product Data sheet for ordering information.
Preparing the Patient
1
Always perform a digital palpation before you insert a rectosigmoidoscope or
endo transducer into the rectum.
2
Give the patient an enema – liquid should be held retained as long as possible
before being expelled.
Preparing the Reusable Water Standoff System
Disinfect
before using
The transducer, water standoff collar, O-ring, rectosigmoidoscope and obturator
must be cleaned and disinfected before use. The water standoff collar, O-ring,
rectosigmoidoscope and obturator can be autoclaved.
NOTE:
For best results, separate the O-ring from the water standoff collar before
autoclaving. O-rings that are no longer watertight should be replaced.
To prepare the reusable water standoff system:
1
Place the two O-rings in the outer grooves as shown in this picture.
2
Use a little water to moisten the silicone O-rings inside the water standoff collar.
3
Gently slide the water standoff collar over the transducer.
A
B
Inner cover
Tape
Outer cover
Water tube
Stopcock
C
The water outlet is just above the
inner groove. Do
NOT
block this
Place the O-rings in
these two grooves
Luer Lock valve
(water inlet)