For Professional Users Only
User Guide
Type 8823
Convex Array Transducer
English BB1740-CJune 2012
Page 1: ...For Professional Users Only User Guide Type 8823 Convex Array Transducer English BB1740 C June 2012...
Page 2: ...rvices As part of our customer satisfaction program we contact a sample of our customers a few months after they receive their orders If you receive an email message from us asking for your feedback w...
Page 3: ...nd Disinfection 6 Starting Scanning 7 Connecting the Transducer 7 Changing Frequency 7 Using a Transducer Cover 7 Using the Transducer Control Button 8 Changing Orientation 8 Puncture and Biopsy Facil...
Page 4: ...4...
Page 5: ...rray Transducer Type 8823 Scanning Plane The scanning plane for 8823 is shown in Figure 2 Figure 2 Scanning plane of 8823 General Information Product specifications for this transducer can be found in...
Page 6: ...s from contact with the patient Turn off the scanner Unplug the scanner from the wall and make sure it cannot be used until it has been checked Do not remove the scanner cover Contact your BK Medical...
Page 7: ...Cover The transducer should be enclosed in a transducer cover or a standard condom See the Product Data sheet for a list of available transducer covers Apply gel to the tip of the transducer This impr...
Page 8: ...g instructions UA1250 The puncture attachment see Figure 3 comprises an attachment bracket an attachment lock screw item A in Figure 4 a needle guide a needle guide lock screw item B in Figure 4 and a...
Page 9: ...ture attachment to the first dot on the scan image puncture line is 52 mm The distance between the dots is 10 mm see Figure 8 for details Figure 4 8823 with UA1250 mounted All parts of the puncture at...
Page 10: ...guide quick release Figure 5 UA1341 and needle guide UA0013 Please consult the Reference Guide that accompanies the single use needle guides for more detailed instructions on how to assemble the punct...
Page 11: ...n image If more than one puncture line is available refer to the applicable scanner user guide for instructions on how to change which one appears Bracket Needle guide WARNING It is essential for the...
Page 12: ...performing RF ablation you must always follow the instructions provided by the manufacturer of the RF ablation equipment Be sure to pay attention to all warnings Do not use excessive force when you in...
Page 13: ...Safety for cleaning instructions Disposal When the transducer is scrapped at the end of its life national rules for the relevant material in each individual land must be followed Within the EU when y...
Page 14: ...14 June 2012 8823User Guide BB1740 C...
Page 15: ......
Page 16: ...LEGAL MANUFACTURER BK Medical ApS Mileparken 34 2730 Herlev Denmark Tel 45 44528100 Fax 45 44528199 Email info bkmed dk...