background image

BT-410F and BT-410A Operation manual

 

P/N:410-ENG-OPM-EUR-R12 

Bistos Co.,Ltd. 

2021,05 

13 

Atmospheric pressure 

70 -106kPa 

 

7 Manufacturer’s declaration on EMC

 

BT-410F and BT-410A needs special precautions regarding EMC (Electromagnetic compatibility) and needs 
to be used according to the EMC information provided in this manual. Wireless communications equipment 
such as wireless home network devices, mobile phones, cordless telephones and their base stations, 
walkie-talkies can affect the head-worn light BT-410F and BT-410A and should be kept at least 1 m away 
from the equipment. 

 

7.1 Electromagnetic emissions 

The head-worn light BT-410F and BT-410A is intended for use in the electromagnetic environment 
specified below. The customer or the user of BT-410F and BT-410A should assure that it is used in such 
an environment.     

Emissions test 

Compliance 

Electromagnetic environment-guidance 

RF emissions 
CISPR 11 

Group 1 

BT-410F and BT-410A is no intentional or controlled RF 
emission for it’s intended performance. Therefore, its RF 
emissions are very low and are not likely cause any 
interference with nearby electronic equipment. 

RF emissions 
CISPR 11 

Class A 

The BT-410F and BT-410A is suitable for use in all 
establishments including domestic, and may be used in 
domestic establishments and those directly connected to the 
public low-voltage power supply network that supplies 
buildings used for domestic purpose. 

Harmonic emission 
IEC61000-3-2 

Class A 

Voltage  fulctuations 
/flicker emissions 
IEC61000-3-3 

Complies 

 

   

WARNING

 

  The  BT-410F  and  BT-410A  should  not  be  used  adjacent  to  or  stacked  with  other 

equipment. If adjacent or stacked use is necessary, the device should be observed to verify 
normal operation in the configuration in which it will be used. 

  Portable  RF  communications  equipment  (including  peripherals  such  as  antenna  cables 

and external antennas) should be used no closer than 30 cm (12 inches) to any part of the 
BT-410F  and  BT-410A,  including  cables  specified  by  the  manufacturer.  Otherwise, 
degradation of the performance of this equipment could result. 

  Use  of  adapter  other  than  those  specified  or  provided  by  the  manufacturer  of  this 

equipment  could  results  in  increased  electromagnetic  emissions  or  decreased 
electromagnetic immunity of this equipment and result in improper operation. 

 

7.2 Electromagnetic immunity 

BT-410F and BT-410A is intended for use in the electromagnetic environment specified below. The 
customer or the user of BT-410F and BT-410A should assure that it is used in such an environment.

 

Immunity test 

Test level 

Compliance level 

Electromagnetic 

environment -guidance 

Electrostatic 
Discharge 

(ESD) 

IEC61000-4-2 

±8 kV 

Contact 

 

±2 kV, ±4 kV, 

±8 kV, ±15 kV 

Air 

±8 kV 

Contact 

 

±2 kV, ±4 kV, 

±8 kV, ±15 kV 

Air 

Floors should be wood, 
concrete or ceramic tile. If 
floors are covered by 
synthetic material, the 
relative humidity should be 
at least 30 %. 

Electrical Fast 
Transient/Burst 
IEC61000-4-4 

±2 kV AC, 

100 kHz PRR 

±2 kV AC, 

100 kHz PRR 

Mains  power  quality  should 
be 

that 

of 

typical 

commercial 

or 

hospital 

environment. 

Surge 
IEC61000-4-5 

±0.5 kV, ±1 kV 

L1 to L2 (DM) 

±0.5kV, ±1kV 

L1 to L2 (DM) 

Mains  power  quality  should 
be 

that 

of 

typical 

Summary of Contents for BT-410A

Page 1: ...BT 410F and BT 410A Head worn light Operation Manual Keep this manual for future reference P N 410 ENG OPM EUR R12...

Page 2: ...e However no responsibility is assumed by Bistos Corporation For its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implicat...

Page 3: ...7 2 1 Product outlook and description 7 2 2 Loupe optional assembly 8 2 3 Astral lamp optional assemble 9 3 Operation 9 3 1 Adjust the headband 9 3 2 Turn on the power 9 3 3 Focusing the light 9 3 4...

Page 4: ...n product label packaging and this manual in order to stand for the information about Stand by Those symbols identify safety information Warning and caution necessarily well know when operating the de...

Page 5: ...old environment to a warm humid location Do not attempt to service the BT 410F and BT 410A Only qualified service personnel should attempt any needed internal servicing 0 3 Battery warnings WARNING Do...

Page 6: ...rief Device Description Product Features Model Configurations 1 2 Brief device description This Head worn light is used for medical examination by fixing it on the head and illuminating with condensed...

Page 7: ...0A has Fixed Focus lens but BT 410F has Variable focus lens BT 410F and BT 410A system consists of the following function and elements Model BT 410F BT 410A Control of Light Intensity X O Control of L...

Page 8: ...ration 2 1 Product outlook and description Fig 2 1 BT 410A outlook and description Headband height adjustable part Headband part Headband size adjustable part Protect band Power On Off button LED indi...

Page 9: ...e shown below Fig 2 3 Loupe Assembly 2 3 Astral lamp optional BT 410A only assembly The astral lamp can be assembled at the bottom of light source Remove the lens cap from the light source and connect...

Page 10: ...e high charging voltage may cause the device malfunction 3 3 Focusing the light BT 410F only You can adjust the focal length between 25 30 cm Adjust the focus for the best light quality before treatme...

Page 11: ...r is unseparated When charging unseparated be sure the device turned off To separate or assemble the battery pack refer below Fig 3 1 Battery pack assembly method Fig 3 2 Battery pack separation metho...

Page 12: ...s during surgery or other cases 4 Maintenance and cleaning 4 1 Checking the light intensity It is recommended that the intensity of the light be checked before each use and at least every six months W...

Page 13: ...h device or you have further questions please contact Bistos customer center 6 Specifications The BT 410F and BT 410A head worn light is classified as follows in accordance with EN IEC 60601 1 Categor...

Page 14: ...ons flicker emissions IEC61000 3 3 Complies WARNING The BT 410F and BT 410A should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the device should be...

Page 15: ...equency magnetic fields should be at levels characteristic of a typical commercial or hospital environment Conducted RF IEC61000 4 6 3 Vrms 150 kHz to 80 MHz 6 Vrms in ISM bands between 0 15 MHz and 8...

Page 16: ...d with the following symbol IEC 60417 Ref no 5140 Non ionizing electromagnetic radiation Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all si...

Page 17: ...t is manufactured and passed through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Framework Act on Consumers noticed by...