Philos DR-T Technical Manual
The following cardiac and technical events initiate a report:
•
Atrial Lead Check < 300 and > 3000 Ohm
•
Ventricular Lead Check < 300 and > 3000 Ohm
•
Ventricular Bursts (Runs)
•
Ventricular Events (Episodes)
•
Low P-Wave Amplitude
1
(< 50% safety margin)
•
Low R-Wave Amplitude
1
(< 50% safety margin)
•
ERI
Activation
•
1
st
Mode Switch/24 hours
W
ARNING
A timely receipt of the corresponding event report cannot be
guaranteed. The receipt is also dependent on whether the
patient was physically situated in the required coverage
range of the patient device at the time the event information
was sent.
N
OTE
:
When ERI mode is reached, this status is transmitted.
Further measurements and transmissions of Home
Monitoring data are no longer possible.
N
OTE
:
The attending physician must notify the BIOTRONIK Service
Center about which of these events he/she wishes to be
informed.
1
Examples: The programmed sensitivity is 1.0 mV.
A) Average of the measured P/R-Wave amplitudes is 2.6 mV. Therefore,
measured value is greater than 100% of the safety margin. Event message is
not triggered.
B) Average of the measured P/R-Wave amplitudes is 1.9 mV. Therefore,
measured value is less than 100%, but greater than 50% of the safety margin.
Event message is not triggered.
C) Average of the measured P/R-Wave amplitudes is 1.4 mV. Therefore,
measured value is smaller than 50% of the safety margin. As a result, an event
message is triggered.