22
Tolerances of Parameter Values
Refractory period A
AUTO
AUTO
—
x
x
Refractory period RV
250 ms
250 ms
300 ms
x
x
x
Refractory period LV
200 ms
200 ms
200 ms
x
Mode switching
ON
ON
—
x
x
Onset criterion
5-out-of 8
5-out-of 8
—
x
x
Resolution criterion
5-out-of 8
5-out-of 8
—
x
x
Intervention rate
160 bpm
160 bpm
—
x
x
Switches to
DDIR
DDIR
—
x
x
The basic rate with
mode switching
+10 bpm
+10 bpm
—
x
x
Rate stabilization with
mode switching
OFF
OFF
—
x
x
PVARP
AUTO (Start
250 ms)
AUTO (Start
250 ms)
—
x
x
PVARP after PVC
400 ms
400 ms
—
x
x
Capture control A
ON
ON
OFF
x
x
x
Capture control RV
ON
ON
OFF
x
x
Capture control LV
ON
ON
OFF
x
Atrial overdrive pacing
OFF
OFF
—
x
x
Vp suppression
OFF
OFF
—
x
IEGM recording (HAR, HVR)
ON
ON
OFF
x
x
x
Home Monitoring
OFF
OFF
OFF
x
x
x
Parameter
Factory
settings
Standard
program
Safe
program
SR DR HF
Parameter
Range of values
Tolerance
Basic rate
30 ... (5) ... 100 ... (10)
... 200 bpm
± 20 ms
Basic interval
1000 ms
± 20 ms
Magnet rate (magnet interval)
90 bpm (664 ms)
± 20 ms
Pulse amplitude
0.2 ... 7.5 V
The greater value of ±50 mV
or +20/-25%
Pulse width
0.1 ... 0.4 ms
± 0.04 ms
0.5 ... 1.0 ms
± 0.10 ms
1.25 ... 1.5 ms
± 0.15 ms
Sensitivity A
EN 45502-2-1 triangle pulse
0.1 ... 0.2 mV
±0,05 mV
0.3 ... 7.5 mV
±20%
RV/LV sensitivity
EN 45502-2-1 triangle pulse
0.5 ... 7.5 mV
±50%
Refractory period
200 ... 500 ms
+ 10/-30 ms
Maximum activity rate
80 ... 180 bpm
± 20 ms
Lead impedance
100 ... 200 Ω
±50 Ω
201 ... 2500 Ω
±25%