Cardiac Airbag Technical Manual
47
[Emergency]
- The [
Emergency
]
function key is always present.
Activating this key produces a screen display that turns
ventricular tachyarrhythmia Detection ON and OFF, activates
emergency bradycardia pacing parameters, or activates a
programmer commanded shock.
of this key
rameter
] window serves for programming parameters for
sensing, detection and therapies. All programming of detection,
[Print]
- This function key appears in the lower right side of the
screen in windows with printing options. Activation
immediately prints the relevant data from the open window.
3.3 Parameter Window
The [
Pa
therapies, pacing are completed from this Parameter screen
(
see Figure
8). Detection and Therapy parameters are
displayed only if VT/VF Detection is enabled. In order to provide
easily understandable parameter displays, the numeric values of
the detection parameters can be shown as an Interval (ms) or as
a Rate (bpm).
Two main function keys are located at the bottom of the
parameter window regardless of the tab selected.
[Interrogate ICD]
- This function key interrogates the ICD and
loads the current programmed parameters into the programmer
memory.
[Transmit]
- Activation of this key transmits the displayed
parameters as a permanent program to the ICD. This key also
appears in the [
Options
]
window as [
Transmit Settings
].