Alternate/Supplemental Tests Using Methylumbelliferone (MUB)
(Optional) As an alternative to using Sodium Fluorescein, Methylumbelliferone (MUB) can
be used to perform the Sensitivity/Linearity tests for the top optics.
Materials
Kits containing the microplates and solutions required by the Liquid Tests
are available for purchase; see
Materials for Conducting Liquid Tests
on
page
.
Manufacturer part numbers are subject to change.
If your reader is equipped with the filter- and monochromator-based
fluorescence systems, the same microplate is used to test both systems.
l
Methylumbelliferone (MUB) (10-mg vial, BTI #98156)
l
Carbonate-Bicarbonate buffer (CBB) capsules (Sigma #3041)
l
100% methanol (BTI #98161)
l
A new, clean 96-well solid black plate, such as Corning Costar #3915 or equivalent. A
Greiner SensoPlate (Mfr. #655892) may also be used.
l
Excitation filter 360/40 installed
l
Emission filter 460/40 nm installed
l
400 nm dichroic mirror installed
l
Deionized or distilled water
l
Various beakers, graduated cylinders, and pipettes
l
95% ethanol (for cleaning clear-bottom plates)
l
Aluminum foil
l
(Optional, but recommended) 0.45-micron filter
l
(Optional) Black polyethylene bag(s) to temporarily store plate(s)
l
Gen5 protocols listed below (as applicable for your reader model) and described in
detail under
starting on page
o
Synergy H1_FI_T_MUB.prt
(filter-based system)
o
Synergy H1_M_FI_T_MUB.prt
(monochromator-based system)
130 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...