Synergy H1 M-LumTest_Harta.prt
The same as the filter-based test, with these exceptions:
Parameter
Setting
READ STEP 1
Optics Type
Luminescence Fiber
Filter Set
<none>
Gain
150
Read Height
1.00 mm
READ STEP 2
Optics Type
Luminescence Fiber
Read wells
F1-G12
Filter Set
<none>
Gain
150
Read Height
1.00 mm
READ STEP 3
Optics Type
Luminescence Fiber
Filter Set
<none>
Gain
80
Read Height
1.00 mm
Results Analysis
The Luminescence Test procedure is described on page
.
1. Determine if the plate’s battery is functioning properly. If A8 > (0.2 * A7), the
battery is good. Otherwise, it requires replacement.
A replacement battery is included with each new and recalibrated Harta
Luminometer Reference Microplate.
2. On the Harta plate’s calibration certificate, locate the NIST measurement for the A2
position. Convert it to
attomoles
: (A2 NIST measurement * 0.02884)
3. Calculate the
signal-to-noise ratio
: (A2–Mean of the buffer cells)/(3 * Standard
deviation of buffer cells)
4. Calculate the
detection limit
: A2 NIST measurement in attomoles/signal-to-noise
ratio
Synergy H1 Operator's Manual
Luminescence Testing | 85