14 | Installation
BioTek Instruments, Inc.
3: Install the MicroFlo’s Components
With the shipping hardware removed, you’re ready to install:
Tubing Cassette
Prime Trough Drain Tube
(Optional) Prime Trough Insert
Fluid supply vessel
Power Cord
(Optional) Serial or USB Cable for communication with computer and/or
BioStack™ Microplate Stacker.
These components are all that is required to run the MicroFlo to process standard
microplates. Additional steps are required to dispense to test tubes and plates with
fewer than 96 wells. Review the section on
Special Plates
in the
Essential Concepts
chapter.
Install the Tubing Cassette
Take a moment to identify the MicroFlo’s parts that are named in the instructions for
installing the tubing cassette.
Figure 3: Dispenser ready for cassette installation.
The Pump Cover must be closed to operate the dispenser.
Pump Cover
Stainless steel plate
Dispense arm
Pump