Getting Started
Page 30 of 42
Epoch 2
Instructions for Use
1771011
Revision
A
BioTek Instruments, Inc.
Run a
Take3
Session
Using the Touchscreen
The
Epoch 2
is preprogrammed with seven
Take3
protocols, available for use after a
Take3
plate has been defined and aligned on the reader.
The protocols measure at 260, 280, and 320
nm
and cannot be edited. For the nucleic
acid protocols, a secondary wavelength of 230
nm
can be selected at run time. For all
seven protocols, a spectral scan may also be selected.
Protocol name
Mass Extinction
coefficient
Secondary Ratio WL
(run time option)
Spectral Scan
(run time option)
dsDNA
50
230
nm
240
–
300
nm
RNA
40
230
nm
240
–
300
nm
ssDNA
33
230
nm
240
–
300
nm
1 Abs@1cm =
1 mg/mL
10
-
260
–
320
nm
BSA
6.7
-
260
–
320
nm
IgG
13.7
-
260
–
320
nm
Lysozyme
26.4
-
260
–
320
nm
Define and Align a
Take3
Plate
Ensure that the glass slide is clean before running the alignment procedure.
Refer to the
Take3
/
Take3
Trio
User Guide
for cleaning instructions.
If you remove or replace the glass slide, perform the Pathlength Calibration
procedure described in the
Take3
/
Take3
Trio
User Guide
and then run the
alignment procedure described below. Update the Pathlength Values on the
touchscreen with the new values generated through the Pathlength Correction
procedure described in the user guide.
The reader can be configured with only one
Take3
plate at a time.
This procedure ensures that the
Take3
slide's microspot positions are captured
correctly in the reader's software. Upon completion, the
Take3
tab appears on the
Main
Menu
and contains the preprogrammed
Take3
protocols.
1.
From the
Main Menu
, tap
Instrument
and select the
Take3
tab.