Test Descriptions
|
85
The second screen shows:
Configuration P/N
Configuration Version
7340203
Version 3.00
Absorbance Plate Test
This test uses BioTek’s Absorbance Test Plate to confirm the Mechanical Alignment,
Accuracy and Linearity, Repeatability, and Channel-to-Channel uniformity of the
ELx808. The Absorbance Plate Test compares the reader’s optical density
measurements and mechanical alignment to NIST-traceable values. To run the test,
you will need a BioTek Absorbance Test Plate with its accompanying data sheet.
The BioTek PN 7260522 Test Plate has a glass filter in location C6, which
is used to check the wavelength accuracy of a monochromator. Although
the ELx808 does not have a monochromator, you must enter values for
the peak wavelength when defining the 7-filter test plate in Gen5
software. Refer to the software help system for more information.
A sample Absorbance Plate Test report is shown on the next page.
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...