12
TurboVap
®
96 Dual User Manual | © Biotage 2021
Evaporate
Evaporate
This section describes how to select different evaporation
modes and how to configure the evaporation settings. It also
describes how to set up and perform an evaporation run and
how to set up a method.
1. Prepare the system as described in "Prepare the System" on
2. Check that the incoming gas supply is turned on and has
sufficient reserve for the evaporation run. Set the incoming
pressure appropriately to achieve desired flow rates, see
"Set the Pressure" on page 11.
3. Select
Mode
from the Mode view, see "Determine the
Appropriate Evaporation Mode" on page 6. Select
Split
to left and right
or
Join left and right
.
4. Configure the evaporation parameters or select method.
5. Close the cabinet doors and press
Start
.
Evaporation Modes
There are three main evaporation modes available;
Manual
mode,
Time
mode and
Method
mode, see Figure 18.
1
Off
The corresponding side is turned off
2
Manual Mode
Evaporate until the user presses the Stop
button
3
Time Mode
Evaporate for a set period of time
4
Method Mode
Evaporate using a predefined method
5
Split to left and right
OR
Join left and right
The left and right cabinets operate
independently
OR
The left and right cabinets operate the
same evaporation run
6
Cancel
Press to cancel mode selection
Figure 18.
The different evaporation modes that can be selected. This
figure shows a dual system setup where left and right cabinets both have
the same manifold configuration (96).
»
Manual Mode:
Evaporate until the
Stop
button is pressed.
»
Time Mode:
Evaporate for a set period of time.
»
Method Mode:
There are several options,
Mode can also be set to:
»
Off:
The corresponding cabinet is turned off, see Figure 19.
»
Split to left and right:
The left and right cabinets
operate independently. The manifold settings do
not have to be synchronized in this mode.
OR
»
Join left and right:
The left and right cabinets
operate the same evaporation run. The manifold
settings must be synchronised in this mode.
»
Cancel:
Mode selection is cancelled.
Figure 19.
In this figure the left side is turned off and the right side is
paused in Manual mode. The left and right sides have different manifolds
(24 and 96 respectively).