7
Technical Specifications
Clog Detection
With an Extrahera GLP software license installed on your system
(sold separately), clog detection is available for load, wash, and
elute operations in the SPE, SLE, and Fil methods.
Supported Solvents
Solvent/Solvent Combination Clog Detection Compatibility
Acetone
Good
Acetonitrile
Excellent
Chloroform
Poor
Dichloromethane
Poor
Diethyl ether
Poor
Dimethylformamide (DMF)
Excellent
Dimethylsulfoxide (DMSO)
Excellent
Ethanol
Excellent
Ethyl acetate
Good
Heptane
Good
Hexane
Poor
Isooctane
Good
Isopropanol
Excellent
Methanol
Excellent
Pentane
Poor
Tetrahydrofuran (THF)
Poor
Toluene
Excellent
Water
Excellent
Methyl tert-butyl ether (MTBE)
Poor
IPA 30%
Excellent
Ammonium hydroxide 5%
Excellent
Formic acid 4.6M
Excellent
DCM/IPA, [90:10]
Poor
DCM/IPA, [95:5]
Poor
DCM/IPA/NH
4
OH [78:20:2]
Poor
DCM/MeOH/NH
4
OH [78:20:2]
Poor
MeOH/NH
4
OH [95:5]
Excellent
Currently Not Supported Consumables
The following Biotage and competitor consumables are not
supported when using clog detection:
»
ISOLUTE
®
PPT+
»
ISOLUTE
®
PLD+
»
ISOLUTE
®
Plate
»
Waters Oasis μElution Plates
»
Tecan NBE and Cerex Columns
»
1 mL Columns, 24 Positions Configuration
»
1 mL Columns, 96 Positions Configuration