background image

 

Appendix G  Calibration 

Frequency 

One of the most common questions 
that we are asked at Biosystems is: 

“How often should I calibrate my gas 
detector?”
 

Sensor Reliability and Accuracy 

Today’s sensors are designed to 
provide years of reliable service.  In 
fact, many sensors are designed so that 
with normal use they will only lose 5% 
of their sensitivity per year or 10% over 
a two-year period.  Given this, it should 
be possible to use a sensor for up to 
two full years without any significant 
loss of sensitivity. 

A lot of sensors indeed do last that long 
with only minimal loss of sensitivity. 
However, there are a number of 
reasons why a sensor may 
unexpectedly lose additional sensitivity 
or even fail to respond to gas. Such 
reasons include desiccation, poisoning, 
physical restriction of airflow, 
overexposure, leakage, and mechanical 
damage due to dropping or immersion. 

Verification of Accuracy 

With so many reasons why a sensor 
can lose sensitivity and given the fact 
that dependable sensors can be key to 
survival in a hazardous environment, 
frequent verification of sensor 
performance is paramount. 

There is only one sure way to verify that 
a sensor can respond to the gas for 
which it is designed. That is to expose it 
to a known concentration of target gas 
and compare the reading with the 
concentration of the gas.  This is 
referred to as a “bump” test. This test is 
very simple and takes only a few 
seconds to accomplish. 

The safest 

course of action is to do a “bump” 
test prior to each day’s use

. It is not 

necessary to make a calibration 
adjustment if the readings are between 
90% and 120% of the expected value. 
As an example, if a CO sensor is 
checked using a gas concentration of 
50 PPM it is not necessary to perform a 
calibration unless the readings are 
either below 45 PPM or above 60 PPM.

 

Lengthening the Intervals 

between Verification of 

Accuracy 

We are often asked whether there are 
any circumstances in which the period 
between accuracy checks may be 
lengthened.  

Biosystems is not the only 
manufacturer to be asked this 
question! One of the professional 
organizations to which Biosystems 
belongs is the Industrial Safety 
Equipment Association (ISEA). The 
“Instrument Products” group of this 
organization has been very active in 
developing a protocol to clarify the 
minimum conditions under which the 
interval between accuracy checks may 
be lengthened.  

A number of leading gas detection 
equipment manufacturers have 
participated in the development of the 
ISEA guidelines concerning calibration 
frequency. Biosystems procedures 
closely follow these guidelines.  

If your operating procedures do not 
permit daily checking of the sensors, 
Biosystems recommends the following 
procedure to establish a safe and 
prudent accuracy check schedule for 
your Biosystems instruments: 

1.  During a period of initial use of at 

least 10 days in the intended 
atmosphere, check the sensor 
response daily to be sure there is 
nothing in the atmosphere that is 
poisoning the sensor(s). The 
period of initial use must be of 
sufficient duration to ensure that 
the sensors are exposed to all 
conditions that might have an 
adverse effect on the sensors.  

2.  If these tests demonstrate that it 

is not necessary to make 
adjustments, the time between 
checks may be lengthened. The 
interval between accuracy 
checking should not exceed 30 
days.  

3.  When the interval has been 

extended the toxic and 
combustible gas sensors should 
be replaced immediately upon 
warranty expiration. This will 
minimize the risk of failure during 
the interval between sensor 
checks.  

4.  The history of the instrument 

response between verifications 
should be kept. Any conditions, 
incidents, experiences, or exposure 
to contaminants that might have an 
adverse effect on the calibration 
state of the sensors should trigger 
immediate re-verification of 
accuracy before further use.  

5.  Any changes in the environment in 

which the instrument is being used, 
or changes in the work that is being 
performed, should trigger a 
resumption of daily checking.  

6.  If there is any doubt at any time as 

to the accuracy of the sensors, 
verify the accuracy of the sensors 
by exposing them to known 
concentration test gas before 
further use.  

Gas detectors used for the detection of 
oxygen deficiencies, flammable gases 
and vapors, or toxic contaminants must 
be maintained and operated properly to 
do the job they were designed to do. 
Always follow the guidelines provided 
by the manufacturer for any gas 
detection equipment you use!  

If there is any doubt regarding your gas 
detector's accuracy, do an accuracy 
check! All it takes is a few moments to 
verify whether or not your instruments 
are safe to use.  

One Button Auto Calibration 

While it is only necessary to do a 
“bump” test to ensure that the sensors 
are working properly, all current 
Biosystems gas detectors offer a one 
button auto calibration feature. This 
feature allows you to calibrate a 
Biosystems gas detector in about the 
same time as it takes to complete a 
“bump” test. The use of automatic 
bump test and calibration stations can 
further simplify the tasks, while 
automatically maintaining records. 

Don't take a chance  

with your life.   

Verify accuracy frequently! 

Biosystems Applications Notes are 
available at the Biosystems website at  

http://www.biosystems.com 

 

Summary of Contents for ZoneGuard

Page 1: ...ZoneGuard Single Channel Fixed Gas Detection System Reference Manual 651 South Main Street Middletown CT 06457 Tel 860 344 1079 800 711 6776 Fax 860 344 1068 Version 2 3 30SEPT2005 p n 13 063...

Page 2: ...F POTENTIALLY DANGEROUS CONDITIONS IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ZoneGuard Operation Manual Version 2 3 Part Number 13 063 Copyright 2005 b...

Page 3: ...__________________________________________8 2 3 REMOTE DETECTOR ASSEMBLIES 8 2 3 1 Standard remote detector housing __________________________________________________8 2 3 2 Explosion proof remote det...

Page 4: ...MANUAL CALIBRATION PROCEDURE 15 3 7 1 Manual Fresh air zero through keypad buttons _______________________________________15 3 7 2 Manual span calibration through keypad buttons _____________________...

Page 5: ...us readings 5 A sensor that cannot be calibrated or is found to be out of tolerance should be replaced immediately An instrument that fails calibration may not be used until testing with known concent...

Page 6: ...e and other components necessary to connect the remote detector assembly with the instrument controller are not included and must be ordered separately The ZoneGuard design includes many user selectab...

Page 7: ...uipped with toxic or combustible gas sensors are provided with two levels of gas alarm Low warning and High danger Toxic and combustible alarms are configured as ascending meaning that they are trigge...

Page 8: ...on proof versions Explosion proof version remote detector assemblies have been submitted to Underwriters Laboratories Inc for Classification per ULC as Explosion Proof for use in Class I Division 1 Gr...

Page 9: ...ate inside the controller case In the case of ZoneGuard systems with integrally mounted sensors a sensor specific amplifier board is mounted onto the relay PCB In the case of ZoneGuard systems with re...

Page 10: ...ic interference EMI the screen drain of the wire should be connected to the ground plate of the detector head long reach glands and or the screen terminals should be used for this purpose Ground loops...

Page 11: ...for one second to turn the ZoneGuard on 2 8 3 Turning the ZoneGuard off Turn the ZoneGuard off by pressing and holding the Mode button until the LCD goes blank When the button is released the LCD wil...

Page 12: ...ings A sensor that cannot be calibrated or is found to be out of tolerance should be replaced immediately An instrument that fails calibration may not be used until testing with known concentration te...

Page 13: ...ment may then be fresh air zero calibrated as discussed in section 3 6 1 3 4 Effect of contaminants on ZoneGuard sensors The atmosphere in which the ZoneGuard monitor is being used can have an effect...

Page 14: ...ses is listed in Appendix E 3 5 Functional bump test Note The functional bump test is only used on instruments equipped with a combustible gas LEL or toxic gas sensor For further details on verifying...

Page 15: ...ndicate that the ZoneGuard CANT ZERO In this case the instrument must be fresh air zeroed using the Cal button on the instrument keypad and procedures discussed in Section 3 7 1 Once the instrument ha...

Page 16: ...ation 3 7 2 Manual span calibration through keypad buttons Span calibration procedures using buttons on the instrument keypad are only necessary when the adjustment necessary to restore accuracy is ou...

Page 17: ...automatic fresh air zero procedure as discussed in Section 3 6 1 Following reinitialization of the instrument or in cases where the ZoneGuard is being used to monitor oxygen levels that are close to 0...

Page 18: ...on to turn the ZoneGuard back on 5 A screen will appear showing the first alarm point to be adjusted the Low warning level alarm set point Use the and keys to raise or lower the Low alarm set point 6...

Page 19: ...includes an Alarm Test function that can be used to automatically ramp readings throughout the full range of the sensor The Alarm Test function is used to verify alarm setpoints as well as the proper...

Page 20: ...tions used during auto calibration adjustment It is necessary to re set the calibration gas concentration stored in instrument memory before making use of a different calibration gas concentration Do...

Page 21: ...pan without saving Set default alarms Turn ZoneGuard on Press ALARM to enter Default Alarm Mode ZoneGuard will display the default alarm settings Press and hold ALARM to restore the factory default al...

Page 22: ...sulfide H2S 25 ppm in nitrogen 58 liters 54 9037 Sulfur dioxide SO2 10 ppm in nitrogen 58 liters 54 9052 Chlorine Cl2 5 ppm in nitrogen 58 liters 54 9055 Nitric oxide NO 30 ppm in nitrogen 58 liters 5...

Page 23: ...ensitivity data The table below lists the cross sensitivity of electrochemical toxic sensors used in Biosystems portable gas detectors to gases other than their target gas Depending on the nature of t...

Page 24: ...Page 23 Appendix F ZoneGuard dimensional drawings electrical schematics and wiring diagrams Dimensional drawing of ZoneGuard controller...

Page 25: ...Page 24 Dimensional drawing of ZoneGuard explosion proof remote detector assembly...

Page 26: ...Page 25 Dimensional drawing of ZoneGuard standard non explosion proof remote detector assembly...

Page 27: ...s procedures closely follow these guidelines If your operating procedures do not permit daily checking of the sensors Biosystems recommends the following procedure to establish a safe and prudent accu...

Page 28: ...nstrument is in service ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Mighty Tox 90 days after ac...

Reviews: