Final Setup
1.
Connect the Liquid Filler accessory to the appropriate port on the Purge Gas Injector
accessory using the cable provided. The Purge Gas Injector accessory should also be
connected to the Ampulmatic-10 base unit.
2.
If using both a Purge Gas Injector Head and a Liquid Filler Injector Head, verify that the
Liquid Filler Injector Head is placed on the Purge Gas Injector Rod first. Position the
Purge Gas Injector Head so that injection occurs at least 2 positions prior to flame sealing.
Position the Liquid Filler Injector Head 1 or 2 positions before the Purge Gas Injector
Head.
3.
If using only the Liquid Filler Injector Head, position it so that injection occurs at least 2
positions prior to flame sealing.
NOTE: Refer to the Purge Gas Accessory Operation and Maintenance Manual for
adjustment of plunger heads.
4.
Fill the carousel with empty ampules and check for correct operation. Visually inspect
ampules before use to discard any broken or distorted ampules. Defective ampules may
cause improper operation. This unit is designed to accept normal variance in glassware;
however, ampules should be of high quality and uniformity for best results.
5.
The Liquid Filler accessory is now set up and ready to use. Turn power on to Ampulmatic-
10 base unit and Purge Gas Injector accessory. Liquid injection will occur automatically.
5-6