Edition 1.-4.03
11
6.
Specifications
The unit is designed for operation in cold rooms, incubators (except CO
2
incubators)
and closed laboratory rooms at ambient temperature from +4°C to +40°C in a non-condens-
ing atmosphere and maximum relative humidity 80% for temperatures up to 31°C decreasing
linearly to 50% relative humidity at 40°C. Operating altitude above sea level is up to 2000 m.
Biosan is committed to a continuous programme of improvement and reserves the
right to alter design and specifications of the equipment without additional notice.
Model
FTA-1
FTA-2i
Aspiration speed
up to 1.7 mL/s
up to 27.8 mL/s
Vacuum regulation
–
Yes
Working vacuum pressure (at sea level)
-500 mbar
-
200 … -800 mbar
Overflow protection sensor
–
Yes, invasive
Overflow protection actions
–
Pump stop, sound and light signals
Trapping flask
1 L, glass
2 L, polypropylene (PP)
Air filtration
Hydrophobic microbiologic air filter 2200/02
1
Filter pore diameter
0.027 μm
Operating voltage
12 V=
Operating current
300 mA
1 A
Power consumption
3.6 W
10.8 W
External power supply
input 100
–240 V~, 50/60 Hz; output 12 V=
Dimensions
160x210x340 mm
185x290x390 mm
Weight, a
ccurate within ±10%
1.7 kg
1.85 kg
1
Eliminates risk of contamination from the trapping flask with bacteria, viruses, and infected particles