This unit can be used in connection with your personal computer running the Blood Pressure Analyzer
Software. The software will allow a capacity of monitoring 80 patients, each with 1000 data (Note:
overuse will lower system efficiency). The memory data can be transferred to the computer by connecting
the monitor via USB cable with your computer.
NOTE:
The USB cable can be purchased at any electronics store. You will need USB cable Mini-B 5 pin.
USB versions 1.0 or 2.0 will work on this device.
System Requirements:
This software is compatible with Microsoft
®
Windows
®
7 / 8 / 10 and supports 32 and 64 bit operating
systems, and Mac OSX 10.10 / 10.11 / 10.12 / 10.13 / 10.14 / 10.15. .
To download the software go to
www.biosmedical.com
. Go to the support page and download BIOS
Diagnostics BP Analyzer Software 3.2.9E for PC and 3.3.5E for MAC (AFIB). Follow the directions to
download the software.
Once installation is complete connect the monitor via USB cable with the PC. Three horizontal bars will
appear on the display and last for 3 seconds.
The bars will then flash to indicate that the connection between computer and device is successfully
made. As long as the cable is plugged in, the bars will keep flashing and the buttons are disabled.
During the connection, the device is completely controlled by the computer. Please refer to the ‘help’ file
in the software for detailed instructions.
5. Error Messages / Malfunctions
If an error occurs during a measurement, a long beep followed by two short beeps is generated and the
LCD display the corresponding error code.
Error
Possible Cause
Remedy
No pulse has been detected or pulse signals
are too weak
Ensure that the cuff is being worn
correctly, and that you have your arm
at the heart level.
Unnatural pressure impulses influence the
measurement result. Reason: The arm was
moved during the measurement.
Avoid unnecessary movement or
talking.
The inflation of the cuff takes too long. The
cuff is not correctly seated.
Ensure that the cuff is being worn
correctly.
Check that the cuff is correctly
connected to the monitor.
The difference between systolic pressure
and diastolic pressure is too far away from
acceptable and reasonable range.
Ensure that the cuff is being worn
correctly and that you have been
inactive for a sufficient time before
making the measurement.