4. Data Memory

At the end of each measurement, this instrument automatically stores the result, including date and time.

4.1 Viewing the stored values

Press the M-button briefly, when the instrument is switched off. The display first shows «

M

» and then a  

value, e.g. «

M17

». This means that there are 17 values in the memory. The instrument then switches to the 

last stored result. Pressing the 

M

-button again displays the previous value. Pressing the 

M

-button repeatedly 

enables you to move from one stored value to another.

4.2 Memory full
 

Pay attention that the maximum memory capacity of 100 is not exceeded. When the memory is full, the old 
values are automatically overwritten with new ones. Values should be evaluated by a doctor before the 
memory capacity is reached – otherwise data will be lost.

4.3 Clearing all values

If you are sure that you want to permanently remove all stored values, hold down the 

M

-button (the instrument 

must have been switched off beforehand) until «

CL

» appears and then release the button. To permanently clear 

the memory, press the M-button while «

CL

» is flashing. Individual values cannot be cleared.

4.4 How not to store a reading

Press the 

ON/OFF

 button while the reading is being displayed. Keep the button pressed until «

M

» is flashing 

and then release it. Confirm by pressing the 

M

-button.

5. Setting the Alarm Function

This instrument allows you to set 2 alarm times at which an alarm signal will then be triggered. This can be a 

useful aid, for instance as a reminder to take medication.

1. To set an alarm time, press the 

TIME

 button (the instrument must have been switched off beforehand) and  

  immediately afterwards the 

M

-button and hold both down until the bell symbol appears in the bottom left  

  of the display. Then release both buttons. The flashing «

1

» in the display indicates that the first alarm time  

  can now be set.
2. Press the time button to set the hours – the hour display flashes and pressing the 

M

-button allows you to set  

    the alarm hour. To confirm, press the 

TIME

 button.

3. The minute display will now flash. The minutes can be set using the 

M

-button. To confirm, press the 

TIME

  

  button again.

4. The bell symbol will now flash. Use the 

M

-button to select whether the alarm time is to be active      or  

  inactive      . To confirm, press the 

TIME

 button. 

5. To set a second alarm time, proceed as above but if the «

1

» flashes, press the 

M

-button to select «

2

» and  

  confirm with the 

TIME

 button.

•  An active alarm time is indicated by the bell symbol in the display.

  The alarm will sound at the set time every day.

•  To switch-off the alarm when it is sounding, press the 

ON/OFF

 button.

• To permanently switch off the alarm, proceed as above and select the      symbol. This will then  

    disappear from the display.

•  The alarm times must be re-entered each time the batteries are replaced.

8

Summary of Contents for PREMIUM A130

Page 1: ...PREMIUM Blood Pressure Monitor Instruction manual...

Page 2: ......

Page 3: ...ce 3 1 Appearance of Patented Arythmia Detection 3 2 Traffic Light Indication 4 Data Memory 4 1 Viewing the stored values 4 2 Memory full 4 3 Clearing all values 4 4 How not to store a reading 5 Setti...

Page 4: ...ircumstances should you alter the dosages of any drugs prescribed by your doctor Depending on physical exertion and condition blood pressure is subject to wide fluctuations as the day progresses You s...

Page 5: ...ing just after you wake up before breakfast and any physical activity and in the absence of the urge to urinate If this is not possible try to take the measurements later in the morning before you sta...

Page 6: ...ht high humidity or in places with a lot of dust Clean the device with a soft dry cloth Do not use gasoline thinner or similar solvents Carefully remove spots on the cuff with a damp cloth and soap Do...

Page 7: ...s Adapter Socket 6 Volume Control 7 Battery Compartment 8 Cuff 9 Cuff Connector 10 M Button Memory 11 Time Button 5 Display 12 Date Time 13 Systolic Value 14 Diastolic Value 15 Pulse 16 Stored Value 1...

Page 8: ...ressed the TIME button the date and time are set and the time is displayed 5 If you want to change the date and TIME press and hold the time button down for approx 3 seconds until the year number star...

Page 9: ...ce the result Note also the explanations on further displays in this booklet 11 When the measurement has finished remove the cuff 12 Switch off the instrument The monitor does switch off automatically...

Page 10: ...pressing the M button 5 Setting the Alarm Function This instrument allows you to set 2 alarm times at which an alarm signal will then be triggered This can be a useful aid for instance as a reminder...

Page 11: ...sibly also set alarm times must be reset the year number therefore flashes automatically after the batteries are replaced 6 3 Which batteries and which procedure Use 4 new long life 1 5V size AA batte...

Page 12: ...r minute Relax for 5 minutes and repeat the measurement LO Pulse too low The pulse is too low less than 40 beats per minute Repeat the measurement 8 Care and Maintanence This instrument may be used on...