BIOPAC Systems NIBP250 User Manual Download Page 3

Noninvasive Blood Pressure Monitoring System 

User's Manual 1. Introduction 

1.1. General Description  

The NIBP250 noninvasive blood pressure monitor is a stand-alone device for measuring the blood pressure of 

small animals. The NIBP250 has a convenient touch screen LCD user interface and includes a built-in Tail Cuff 

Sensor pump that automatically inflates the cuff to occlude the blood vessels in the tail of a rat or similar small 

animal. Once the pump reaches maximum inflation point the cuff slowly deflates, providing a linear drop in 

pressure. Using the simple touch screen interface, the blood pressure points are easily marked for analysis. The 

NIBP250 returns marked values for systolic and diastolic blood pressures and calculated values for mean blood 

pressure and heart rate. The NIBP250 does not require a data acquisition system or computer to analyze the 

data, but can interface with existing BIOPAC data acquisition platforms and third-party systems when combining 

other physiological signals. 

1.2. Theory of Measurement  

The NIBP250 noninvasive blood pressure system uses an adapted version of the well-known ‘Oscillometric 

Technique

1

. Oscillometric Technique was first demonstrated in 1876 and involves the observation of oscillations in 

the sphygmomanometer cuff pressure, which are in turn caused by the oscillations of blood flow, i.e., the pulse. 

An electronic version of this method is sometimes used in long-term measurements and general practice. This 

method relies on a sphygmomanometer cuff with an electronic pressure sensor (transducer) to observe cuff 

pressure oscillations. By use of this type of sensor, the NIBP250 automatically acquires BP data, and controls the 

automatic inflation and deflation of the cuff.  
The cuff is initially inflated to a pressure in excess of the systolic arterial pressure and then reduced over a period 

of about 30 seconds to just below diastolic pressure. When blood flow is nil (cuff pressure exceeding systolic 

pressure) or unimpeded (cuff pressure below diastolic pressure), cuff pressure will be essentially constant. When 

blood flow is present but restricted, the cuff pressure (monitored by the pressure sensor,) will vary periodically in 

synchrony with the cyclic arterial expansion and contraction, i.e., it will oscillate. The systolic and diastolic 

pressures are computed using an algorithm, and the computed results displayed. 
For human use, blood pressure cuffs are snugly wrapped around the upper arm at roughly heart level. The blood 

flow rate in the upper arm is measured to determine whether the flow is impeded (systolic) or unimpeded 

(diastolic). For use in small laboratory animals, the cuff is placed around the tail. However, the blood flow rate in 

the tail is normally too low for the oscillations to be detected with a sphygmomanometer cuff. The NIBP250 

system accurately measures systolic and diastolic pressure levels using a Tail Cuff Infrared Pulse Sensor to 

detect blood flow pulsations. To enhance measurement accuracy, it’s advisable to gradually heat the tail of the 

animal using the optional Tail Heating Unit.  

1.3. What’s Included in System 

Each NIBP250 system includes:  

  1 - amplifier unit (switching power supply works for 110 V or 220 V)  

  1 - cuff/sensor  

o

  size 11 mm is included, but can be specified as 9.5 mm or 13 mm when ordering  

o

  optional MRI-compatible cuff/sensors available as add-ons: 9.5 mm11 mm, or 13 mm  

  1 - restrainer (size medium is included, but can be specified as small or large when ordering)  

  1 - Pulse cable to MP System or third-party A/D hardware  

  1 - Pressure cable to MP System or third-party A/D hardware  

 

                                                 

1  Pickering, TG; Hall, JE; Appel, LJ et al. (2005). "Recommendations for blood pressure measurement in humans 

and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from 
the Subcommittee of Professional and Public Education of the American Heart Association Council on High 

Blood Pressure Research". Hypertension 45 (5): 142–61.  

[email protected]

 

Page 3 of 18 

Summary of Contents for NIBP250

Page 1: ...NIBP250 User s Guide Noninvasive Blood Pressure Measurement System 42 Aero Camino Goleta CA 93117 Telephone 805 685 0066 Facsimile 805 685 0067 support biopac com www biopac com...

Page 2: ...rs 13 4 Using the NIBP250 with a BIOPAC MP Acquisition System 14 4 1 Connections 14 4 2 Software Setup in AcqKnowledge 14 5 Technical Specifications 15 6 Troubleshooting 16 6 1 Tail Pulse signals are...

Page 3: ...low diastolic pressure When blood flow is nil cuff pressure exceeding systolic pressure or unimpeded cuff pressure below diastolic pressure cuff pressure will be essentially constant When blood flow i...

Page 4: ...to reach the selected temperature 7 Remove the RESTRAINER from the Animal Heating Chamber 8 Connect the Tail Cuff Sensor to the animal s tail 9 Verify that the sensor fits the tail correctly The sens...

Page 5: ...7 USB Memory Slot 3 Pulse Signal Analog Output 8 SD Card Slot 4 Pressure Signal Analog Output 9 Tail Sensor Input 5 Trigger Input 10 Tail Cuff Tubing Port 2 3 Sensor Figure 3 shows Tail Cuff Sensor an...

Page 6: ...ttons Button Functionality Button Functionality Moves the cursor to the left or right by one step If Peak by peak is enabled the cursor moves to the previous or next peak Saves the last measurement Se...

Page 7: ...option is selected the value of the peak is displayed If unchecked the value of the measurement sample is displayed To record systolic and diastolic points move the cursor to the desired axis position...

Page 8: ...6 Menu window 2 5 1 General Tab The General Tab see Figure 7 is where the NIBP200 date and time is set or adjusted Press the Set button to display the Date Time Properties see Figure 8 Figure 7 Genera...

Page 9: ...2 5 3 Calibration In the Calibration Tab see Figure 10 the pressure sensor calibration can be checked and recalibrated if necessary Figure 10 Calibration Tab Settings The Calibration Tab displays the...

Page 10: ...ns the leakage valve to release the pressure After the pressure is fully released the acquisition stops 3 3 Analysis Following the acquisition the unit automatically detects the pulse peaks after whic...

Page 11: ...the pulse with the highest peak and then press the Diastolic button The system will record and display the diastolic blood pressure value You may change your cursor peak positions at anytime during t...

Page 12: ...pressure value Heart rate value Description text Data sample period in seconds Number of data lines Pressure value Pulse value at time 0 s Pressure value Pulse value at time 0 004s Pressure value Pul...

Page 13: ...pressure value shown by the manometer and the Current Pressure value on the NIBP250 s screen 9 If the two values match the calibration is validated If not recalibration is necessary 3 9 2 Recalibrati...

Page 14: ...with BIOPAC MP36 35 Data Acquisition Unit with BSL PRO software use the CBL35 PRE for Step 1 and CBL35 PLS for Step 2 NOTE BIOPAC recommends connecting the CBL150 PRE cable to the first channel A1 so...

Page 15: ...e Sensitivity 0 1 mmHg Pressure Signal Output 300 mmHg 3 Volt DC Pulse Gain Levels x1 x2 x4 x5 x8 x16 x32 adjustable Pulse Signal Output 0 5 Volt DC LCD Display 7 800x480 TFT User Interface Resistive...

Page 16: ...sor in a different location on the tail The optimal location is between the mid point of tail and base of tail spinal column 6 2 Compressor runs continuously Immediately Turn off the NIBP250 system Re...

Page 17: ...1 mm RXTCUFSENSOR13 Tail Cuff Sensor 13 mm RXCUFSEN9 5 MRI Tail Cuff Sensor for MRI 9 5 mm RXCUFSEN11 MRI Tail Cuff Sensor for MRI 11 mm RXCUFSEN13 MRI Tail Cuff Sensor for MRI 13 mm RXRESTRAINER S Re...

Page 18: ...Representative Provide the following information Your Name Unit Serial Number Your Institution Sensor Size Address Describe the problem Telephone Number Facsimile Number E mail Address You will be ad...

Reviews: