FOTS100 User Manual 

 

© Opsens Inc. 

IMP0023 Pico-M Manual rev 2.1 test.doc 

1. Quick start 

1.1  FOTS100 Powering 

Connect the plug-in wall or tabletop power supply to a power outlet. Verify that the power 

supply complies with the voltage and frequency outlet. Connect the power cable to the 

FOTS100 power connector (Figure 1). 
 

 

Figure 1 : Electrical connectors 

 

 

1.2  Sensor connection 

The FOTS100  is compatible with all TSD180 series Opsens’ GaAs-type of sensors or 

transducers for measuring temperature namely the OTG-M  series.  The  fiber-optic 

connector provided with the OTG-M  sensors  is  a  ST-type connector.  Remove the 

protective cap of the FOTS100  ST mating and engage the sensor connector  with the 

orientation key properly oriented  (Figure  2).  It is a good practice to clean the sensor 

connector prior to connect it to the signal conditioner. See the TSD180/181 Fiber-Optic 

Cleaning Guide for further information on how to clean fiber-optic connectors.  

NOTE: always replace the protective dust  cap on the mating when there  is no 

sensor connected.  Always replace the protective  dust  cap  on the sensor fiber-

optic connector when not in used 

RS-232 Connector 

 

Wall Supply connector 

 

±5V Analog output 

Summary of Contents for Opsens FOTS100

Page 1: ...FOTS100 User Manual Opsens Inc IMP0023 Pico M Manual rev 2 1 test doc 1 FOTS100 User Manual BIOPAC Systems Inc 42 Aero Camino Goleta CA 93117 Tel 805 685 0066 Fax 805 685 0067 WWW BIOPAC COM...

Page 2: ...le for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material This warranty does not apply to the transducers sold for...

Page 3: ...OSION PROOF Installation of this instrument in an area requiring devices rated as intrinsically safe is not recommended WARNING VOLTAGE SUPPLY Use only the wall plug in power supply delivered with you...

Page 4: ...1 2 Menu Button 7 2 1 3 Left Right Arrows 7 2 1 4 Up Down Arrows 7 2 1 5 Confirmation button 7 2 1 6 Log button 7 2 2 System Setting 7 2 2 1 Averaging mode 7 2 2 2 Analog output 8 2 2 3 Units 8 2 3 S...

Page 5: ...namely the OTG M series The fiber optic connector provided with the OTG M sensors is a ST type connector Remove the protective cap of the FOTS100 ST mating and engage the sensor connector with the ori...

Page 6: ...0 is ready to measure temperature The display shows the temperature measurements with a refreshing rate of three readings per second The measurement readings are also available at the analog output an...

Page 7: ...s Up Down arrows allow navigating 1 between menu items of a given hierarchical level or 2 it allows changing a value being entered 2 1 5 Confirmation button Confirmation button permits confirming 1 a...

Page 8: ...Temperature Voltage output x 10 C V 5 C The default value of the scale factor is 50 C V or its equivalent in F and the default value of the offset is 0 C or its equivalent in F NOTE The analog output...

Page 9: ...r is connected to the unit take note of the diagnostic parameters and contact BIOPAC technical support During a No Signal condition the analog output and the serial ports output constant values as fol...

Page 10: ...ens Inc IMP0023 Pico M Manual rev 2 1 test doc 10 2 4 Setting the averaging measurements Menus levels Figure 3 Averaging measurements Averaging twenty sequential measurements are averaged for obtainin...

Page 11: ...nual Opsens Inc IMP0023 Pico M Manual rev 2 1 test doc 11 2 5 Setting the analog output Menus levels Figure 4 Setting up the analog output Both the scale factor and offset can be setup according to us...

Page 12: ...al Opsens Inc IMP0023 Pico M Manual rev 2 1 test doc 12 2 6 Setting the temperature units Menus levels Figure 5 Setting the temperature units The temperature units can be set to degree Celsius or to d...

Page 13: ...igure 6 Diagnostic The user can look through a variety of diagnostic parameters for diagnosing potential problem with the FOTS100 or the fiber optic sensor The available diagnostic parameters are as f...

Page 14: ...commands available for the FOTS100 But for ease of operation Opsens provides its own control software called SoftSens which gives access to all the functionalities of the FOTS100 conditioner See Soft...

Page 15: ...available Resolution 0 C Sampling rate 50 Hz 20 ms Output interface Display 5 Volts and RS 232 standard Communication protocol SCPI default Input power 9 to 24 VDC positive is center pin 12 V AC DC wa...

Reviews: