BIONIME RIGHTEST GT200 Manual Download Page 2

Storage and Handling

- Store the strips in the original capped vial at temperatures between 4°C - 30°C (39°F - 86°F) and relative humidity 

below 90%. Do not freeze.

- Replace vial cap immediately and close lid tightly after removing a test strip.  If the strip is exposed to the air too 

long, it will absorb the moisture and cause inaccurate test results.

- When you open a new vial of test strips please write the opening date on the label. Use test strips within 12 months 

after first opening or until the expiration date printed on the label (whichever comes first).

Quality Control

Please refer to the Quality Control section of the RIGHTEST GT200 Blood Glucose Monitoring System User’s Manual.

Troubleshooting and Customer Service

For more information on error messages and troubleshooting, please refer to the Error Messages and 
Troubleshooting section of the RIGHTEST GT200 User’s Manual.

If you have any questions or in case of problems with the RIGHTEST products, please contact local Bionime distributor 
or email to [email protected] .

NOTE
- Suggest not to use this meter close to source of strong electromagnetic radiation, to avoid interference with 

proper operation.

- Suggest to keep meter free of dust, water or any liquid.

Additional Information for Healthcare Professionals

(2)

Detection Principle

The glucose Dehydrogenase and potassium ferricyanide in the strip react with the glucose in the sample to produce an 
electrical current which is proportional to the amount of glucose in the sample. The meter measures the current and 
converts it to the corresponding glucose concentration.

Performance Characteristics

Measurement Range
The measurement range of the 

 is 10 to 600 mg/dL 

.

RIGHTEST GT200 Blood Glucose Monitoring System

(0.6 to 33.3 mmol/L)

Precision

The precision was evaluated with (i) venous whole blood sample (ii) 5 different glucose concentrations in 10 days by 10 
meters. 

P-01

P-02

P-03

P-04

P-05

300

42.2 (2.3)
1.1 (0.06)

2.6%

300

98.2 (5.5)
1.8 (0.10)

1.9%

300

131.6 (7.3)

2.8 (0.16)

2.1%

300

233.5 (13.0)

4.5 (0.25)

1.9%

300

352.6 (19.6)

5.8 (0.32)

1.6%

CS-L

CS-N

CS-H

300

57.5 (3.2)
1.8 (0.10)

3.1%

300

108.4 (6.0)

2.5 (0.14)

2.3%

300

272.3 (15.1)

6.3 (0.35)

2.3%

Accuracy

The accuracy of RIGHTEST GT200 Blood Glucose Monitoring System was tested by comparing fingertip whole blood 
(plasma equivalent) glucose values measured by RIGHTEST GT200 with plasma glucose values obtained from a YSI 2300 
(Calibrated with NIST 917c) reference instrument.
The result is shown in the tables below.

*Acceptance criteria in ISO 15197 : 2013 are that 95% of all differences in glucose values should be within ± 15 mg/dL 
(0.83 mmol/L) at glucose concentrations < 100 mg/dL (5.55 mmol/L), and within ± 15 % at glucose concentrations 

≧ 

100 mg/dL (5.55 mmol/L).

Note:

 For glucose concentrations < 100 mg/dL (5.55 mmol/L), difference values are expressed in mg/dL (mmol/L), and 

for glucose concentrations 

100 mg/dL (5.55 mmol/L), difference values are compared inpercentage.

Lay User Evaluation

A total of 104 users were enrolled. Each user tested their fingertip blood samples with RIGHTEST GT200 test strip and 
meter. Then the professional collected blood samples were centrifuged immediately after collection to obtain plasma. 
Analyze the plasma by the lab instrument (YSI 2300 analyzer). 100% of the BGMS values were within 15% of YSI values at 
glucose concentrations 

100 mg/dL (5.55 mmol/L) and within± 15 mg/dL (0.83 mmol/L) at glucose concentration < 100 

mg/dL (5.55 mmol/L).

Hematocrit(Hct)

Hematocrit(Hct) should be between 20%-70% when blood glucose 

 200 mg/dL (11.1 mmol/L), Hct 20%-60% when blood 

glucose > 200 mg/dL (11.1 mmol/L). If you do not know your hematocrit , ask your healthcare professional.

Interferences

26 toxic amount tested substances (Acetaminophen, Ascorbic acid, Dopamine, EDTA, Gentisic acid, Heparin, Ibuprofen, 
L-dopa,  Methyldopa,  Pralidoxime  iodide,  Salicylic  Acid,  Tetracycline,  Tolazamide,  Tolbutamide,  Bilirubin,  Cholesterol, 
Creatinine,  Glutathione,  Haemoglobin,  Triglycerides,  Uric  acid,  Maltose,  Xylose,  Galactose,  Lactose,  Icodextrin)  in  two 
blood sample concentrations.
Substance and possible level may interfere the glucose measurement:
Ascorbic acid 

≧ 

5 mg/dL (0.28 mmo/L)                

Uric acid 

≧ 

20 mg/dL (1.19 mmol/L)                 

Xylose 

≧ 

20 mg/dL (1.33 mmol/L)

Reagents

Each Blood Glucose Test Strip contains the following reagents:     
1. FAD-Glucose dehydrogenase              12.4 %
2. Potassium Ferricyanide                    

 

 49.6 %

3. Non-reactive Ingredients

         

     38.0 %

References

1) Diabetes Information - American Association for Clinical Chemistry (AACC)

Electronic Version

Retrieved Mar. 02, 

2018 form www.labtestsonline.org/understanding/analytes/glucose/test.html

2) In Vitro Diagnostics in Diabetes: Meeting the Challenge. Clinical Chemistry 45:9, 1596-1601 (1999).

(1) Total test numbers
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV

(1) Total test numbers
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV

(i) Venous whole blood sample: 

(ii) Control solution:

Glucose levels

Glucose levels

Bionime GmbH
Tramstrasse 16
9442 Berneck
Switzerland
E-mail: [email protected]

BIONIME CORPORATION

No. 100, Sec.2, Daqing St., South Dist., 
Taichung City 40242, Taiwan
Tel:  +886 800-241-688
http: //www.bionime.com            
E-mail: [email protected]

0197

101-3GS750-

XXX

EN

Rev. Date:2019-04

Store  between  tempe rature  4

  and  30

  ( 39  and  86    ) 

°

C

°

C

°

F

 

°

F

For 

in vitro

 diagnostic use

Manufacturer

LOT

IVD

Lot number

Expiry date

EC Representive

Do not reuse

CE-mark (with No. of notified body)

Biological risks

Consult the instruction for use

Capillary samples collected
from different testing sites

Fingertip

Palm

Forearm

The percent (and number) of samples for which the difference between the RIGHTEST 
GT200 Blood Glucose Monitoring Meter value (Alternate site) and the YSI 2300 value 
were within the following intervals.

Within ± 5%

318/420 (75.7%)

300/420 (71.4%)

296/420 (70.5%)

Within ± 10%

411/420 (97.9%)

412/420 (98.1%)

395/420 (94.0%)

Within ± 15%

420/420(100.0%)

420/420(100.0%)
420/420(100.0%)

Table 3: represents samples for glucose concentrations 

100 mg/dL (5.55 mmol/L).

Table 1: Accuracy basic information

Test range in mg/dL (mmol/L)

Within ±15mg/dL 
(0.83 mmol/L) or within ±15 %

25-599 (1.39-33.28)

624/624 (100.0%)

26-599 (1.44-33.28)

624/624 (100.0%)

24-599 (1.33-33.28)

624/624 (100.0%)

Fingertip

Palm

Forearm

Capillary samples 
collected from different 
testing sites

Fingertip

Palm

Forearm

The percent (and number) of samples for which the variation between the RIGHTEST 
GT200 Blood  Glucose Meter value (Alternate site) and the YSI 2300 value were within the 
following intervals.

Within ± 5 mg/dL (0.28 mmol/L)

150/204 (73.5%)

145/204 (71.1%)

140/204 (68.6%)

Within ± 10 mg/dL (0.56 mmol/L) 

201/204 (98.5%)

200/204 (98.0%)

192/204 (94.1%)

Within ± 15 mg/dL (0.83 mmol/L)

204/204 (100.0%)

204/204 (100.0%)

204/204 (100.0%)

Table 2: represents samples for glucose concentrations< 100 mg/dL (5.55 mmol/L).

Reviews: