45
44
Understanding Control Test Results
Recalling Test Results
The
Blood Glucose Meter
is able to automatically store a maximum of 500 test
results with time and date. If your meter has stored 500 results, the newest test result will replace
the oldest one.
To recall your test memory, start the meter without a test strip inserted.
TM
Rightest
GM782
1. Press the Main button to switch from the Time Mode to Memory Mode. You
will see the “mem” icon in the upper left corner of the display. The latest test
result will appear with the number “1”.
2. Use either the Left or Right button on the side to review all previous results
with date and time. Results display from the most recent (Sequence no. “1”)
to the oldest (Sequence no. “500”) in the lower right hand corner of the
screen.
3. The Quality Control Test result can be recalled from the saved data. When
" appears next to the data, the test was performed using the
Control Solution GC700. The control solution test result will not be used when
recalling test average results.
TM
"
Rightest
Possible reasons your Control Solution results are out of the range :
- Your
Control solution GC700 is expired or was first opened more than 3 months ago.
- Your
Blood Glucose Test Strip GS700 has expired.
- You left the cap of the Blood Glucose Test Strip vial or the control solution off for a period of time.
- You did not perform the test procedure correctly.
- The
Blood Glucose Meter or
Blood Glucose Test Strip GS700 have
malfunctioned.
If
Control Solution GC700 results are out of range, your
Blood Glucose
Monitoring System GM782 may not be working properly. Repeat the Quality Control Test. If your
control solution results are still out of range, do not use
Blood Glucose Meter
to test your blood glucose. Please contact
TM
Rightest
TM
Rightest
TM
TM
Rightest
Rightest
TM
TM
Rightest
Rightest
TM
Rightest
customer service.
GM782
GM782
88~120 mg/dL
250~340 mg/dL
Normal
High
40~60 mg/dL
Low
Your control solution test results should fall within the control solution range printed on the test strip
vial label. If the results are within the range, the
Blood Glucose Monitoring System
GM782 is working correctly.
TM
Rightest
Example of control solution range printed on your test strip vial label