BIONIME rightest GM700S User Manual Download Page 26

49

48

Er1 -

 

Er2 -

 The m

Er3 - 

The s

The inserted test strip has been used or damaged. Please use a new test 
strip.

eter has malfunctioned. Do a Quality Control Test or reinstall the 

batteries to check if the meter works properly.

ignal transmission is disrupted. Repeat the test using a new test strip.

Repeat the test using a new test 

strip.

Er5 - 

An issue calibrating the Meter has occurred. Please follow the steps below:

  (1) Remove the test strip from the meter.
  (2) Turn off the meter (press and hold the Main button for 3 seconds).
  (3) Press the main button to turn ON the meter.

  If after turning on the Meter, you do not see Er5, your meter is functioning 

properly and able to perform a test.

If one of the above error messages still appears, please contact Bionime  
Customer Service.

Er4 - 

The blood sample volume is insufficient. 

Error Messages and Troubleshooting

Meter Malfunction

If the 

 Blood Glucose Meter

 will not turn on, please follow the steps below:

1. Open the battery cover and remove the batteries.
2. Wait for 5 minutes and reinsert the batteries as described as instructed on page 16, (Meter 

Activation and Changing the Battery).

The meter should work normally after finishing the above steps. If the meter still does not work, 
please contact your local 

 customer service.

TM

 

Rightest

GM700S

 

Bionime

Strip Error

"           "

TM

Rightest

"          "

"          "

TM

Rightest

TM

Rightest

Sampling Error

Please do not apply the blood to the sample port of the test strip before the 
meter displays "      ". Discard the test strip If the meter shows "          ".

1. When the test strip is inserted incorrectly, the meter will beep 4 times while the 

 symbol flashes. Do not apply a sample on the test strip. Please re-

insert the unused test strip correctly (refer to “Handling the 

 Blood 

Glucose Test Strip GS700“ on Page 23 for more information).

2. If, after performing the required steps, the 

 and flashing 

 

symbols appear on the screen, it means the 

 Blood Glucose Test 

Strip GS700 has been inserted incorrectly more than twice. Please, re-insert 
the unused test strip again after reviewing the steps on how to insert a test 
strip correctly on p.23 or call Customer Service for support. This error message 
can also indicate that you may have used the wrong test strips. Please check 
the test strip vial to ensure you are using the 

 Blood Glucose Test 

Strip GS700. 

Battery Error

1. The "         " and "          " symbols appear and the meter will not function when 

the battery is low. Change the batteries immediately.

Error Messages and Troubleshooting

Summary of Contents for rightest GM700S

Page 1: ...Berneck Switzerland E mail info bionime ch GM700S 0197 BIONIME CORPORATION No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E...

Page 2: ......

Page 3: ...by Bionime Corporation If you have any questions or concerns please contact your local Customer Service or call us at 886 0800 371 688 GMT 08 00 08 30 17 30 Bionime Customer Thank you for selecting th...

Page 4: ...ay give inaccurate results TM The Rightest Blood Glucose Monitoring System GM700S is intended for in vitro diagnostic use TM only The Rightest Blood Glucose test results using fresh capillary whole bl...

Page 5: ...Alternative Site Testing View Window Appearance Understanding Test Results and Messages TM Rightest GM700S TM Rightest GM700S TM Rightest GM700S TM Rightest GS700 TM Rightest GS700 8 10 14 16 18 22 2...

Page 6: ...ucose Meter with 1 CR2032 batteries installed 6 Blood Glucose Test strips GS700 10 25 pcs 7 Control Solution GC700 Normal Level is provided in your starter kit 8 Lancing Device 9 Clear Cap 10 Disposab...

Page 7: ...The Blood Glucose Meter TM Rightest GM700S Right Button Battery Cover Meter Battery The Blood Glucose Meter TM Rightest GM700S Left Button Main Button Display Window Test Strip Port 2D barcode with s...

Page 8: ...esult Test result Warns when the batteries are low or must be replaced Indicates when to apply the blood sample Appears after you insert test strip into meter Appears when an error occurs Current date...

Page 9: ...est Strip GS700 record the date on the vial TM Rightest Blood Glucose Test Strip GS700 TM If the Rightest Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature please...

Page 10: ...ease follow the local regulation to properly discard a used battery 4 The Blood Glucose Meter 5 Press any button to exit the self test and enter Setting Mode TM Rightest replacing the battery all symb...

Page 11: ...onds During this time the screen will go blank until you hear a beep After the beep the meter will turn on into the Setting Mode The display screen will show setting data GM700S GM700S Setting Up Your...

Page 12: ...change any settings during Meter Set up for over 2 minutes the Blood Glucose Meter will leave setting mode and power off automatically TM Rightest GM700S Setting Up Your Meter Setting the Date Time a...

Page 13: ...no strip is inserted GM700S GM700S GM700S Turning On Off the Meter Handling the Blood Glucose Test Strip TM Rightest GS700 How to handle the ood Glucose Test Strip Inserting the Blood Glucose Test St...

Page 14: ...f the test strip port Please follow your local regulations and discard used strips properly TM Rightest GS700 1 2 3 Getting Ready for Testing Before performing a blood glucose test prepare the items b...

Page 15: ...ed skin 1 2 3 4 5 6 Performing a Blood Glucose Test 7 Hold the base in one hand and pull on the plunger with the other hand The device will be cocked Release the plunger and it will automatically move...

Page 16: ...repeat with a new Blood Glucose Test Strip GS700 15 Countdown Mode will begin on the display window After 5 seconds your test result will appear TM Rightest 13 15 Performing a Blood Glucose Test 14b 1...

Page 17: ...uld wash hands thoroughly with soap and water after handling the meter lancing device and test strips TM Rightest TM TM Rightest Rightest GM700S Setting Markers According to actual life status you can...

Page 18: ...ratory test system that is known to be well maintained and monitored by a healthcare provider use the clear TM cap provided with the Rightest Lancing Device GD500 when testing Alternative Site Testing...

Page 19: ...provider for appropriate disposal of used test strips and lancets TM Rightest TM Rightest TM TM Rightest Rightest GM700S Make sure your blood sample covers the whole area of the view window to get an...

Page 20: ...t TM Rightest Control Solution Range Example of Control Solution Range printed on your test strip vial label Control Solution Range mg dL Low 28 48 Normal 107 145 High 315 426 NOTE Normal of Control S...

Page 21: ...ate 2014 06 When should a Quality Control Test Be Performed To ensure that your Blood Glucose Meter and Blood Glucose Test Strip are working properly To confirm that you are following the correct test...

Page 22: ...trol solution This could contaminate the meter via the test strip port Keep the test strip port clean and dry Clean immediately if the test strip port is stained or is overly exposed to moisture Do no...

Page 23: ...lts TM Rightest Possible reasons your Control Solution results are out of the range Your Control solution GC700 is expired or was first opened more than 3 months ago Your Blood Glucose Test Strip GS70...

Page 24: ...lts control solution test results and test results under abnormal temperature 43 F 6 C 111 F 44 C will be excluded from average calculations The Average function requires that the correct time and dat...

Page 25: ...ort is stained with blood control solution or any liquid please use a dry tissue or alcohol swab to clean it immediately Do not use anything wet to clean it Perform a quality control test to ensure th...

Page 26: ...The meter should work normally after finishing the above steps If the meter still does not work please contact your local customer service TM Rightest GM700S Bionime Strip Error TM Rightest TM Rightes...

Page 27: ...other interference keep meter out of electromagnetic radiation sources TM Rightest Inaccurate test results may be obtained at altitudes greater than 10 000 feet 3048 meters above sea level Xylose 20 m...