4
5
System Components
BATTERY PACK AND BATTERY CHARGER
The
Biomet
®
SpinalPak
®
Non-invasive Spine Fusion Stimulator System includes two rechargeable battery packs.
Upon receipt, it is recommended that the second battery pack be immediately placed into the charger cradle and fully
charged. In the meantime the first battery pack may be used to begin your treatment.
Note: The first battery pack may not provide a 24-hour treatment initially. It is recommended that the patient keep one
battery pack in the battery charger cradle to assure that it is fully charged and ready, and the other inserted into the
Biomet
®
SpinalPak
®
Non-invasive Spine Fusion Stimulator System. This will ensure continuous treatment prescribed by
the prescribing physician.
The battery charger and cradle are designed to recharge the
Biomet
®
SpinalPak
®
Non-invasive Spine Fusion Stimulator
System battery packs only. Two LED (light emitting diode) lights monitor and indicate the operational status of the battery
charger. The A/C Power indicator light is located on the A/C Adaptor. The charging status indicator light is located on the
charger cradle. The following table lists and describes the operational functions of the LED lights:
Electrodes
There are two types of electrodes that are packaged with the Biomet
®
SpinalPak
®
Non-invasive Spine
Fusion Stimulator System assembly: 72R and LT-4500. The 72R electrodes have green writing on their
packaging. The LT-4500 electrodes have black writing on their packaging. The 72R electrodes have a
hydrogel that is stickier than the LT-4500 electrode hydrogel. The patient can use whichever electrodes
best suit their skin type.
Electrode Covers
The electrode covers are water resistant and are intended to enhance electrode security to the skin, if
needed, or for showering with the electrodes attached, if desired.
Device Holster
The device holster is designed to securely hold the Biomet
®
SpinalPak
®
Non-invasive Spine Fusion
Stimulator System in place. It has a clip on the back which allows the patient to wear the device on
their waistband or belt.
Lead Wires
Two different length lead wires are included with the Biomet
®
SpinalPak
®
Non-invasive Spine Fusion
Stimulator System. The patient should choose the lead wire that best accommodates their needs for
where they would like to wear the control unit.
Following are possible error conditions and possible resolutions.
Status
A/C Power
Indicator Light
Charging Status Indicator
Light on Cradle
No battery pack inserted
(idle) on A/C powered
battery charger
Solid green
Off
Battery pack in charging state
Solid green
Solid orange
Fully charged battery pack
Solid green
Solid green
A/C power deficiency
Off
Off
Error
Solid green
Off
Error Conditions (flashes orange)
Possible Resolutions
Battery pack not properly connected
to the charger
Remove and re-install the battery pack to
ensure a complete connection to the charger
Battery temperature is too low or high
Normal operating temperature is
5°C to 38°C [41°F - 100°F]
Battery voltage is too low
Call Biomet for a new battery pack
Summary of Contents for SpinalPak