If the audible alarm and the orange light flashes with the indicator 

 icon blinking, the battery charge is low. Install a fully 

charged battery pack.

If the audible alarm is beeping, the orange light is blinking, and 
the 

 icon is displayed, the continuity of the circuit has been 

interrupted. First, check that the electrodes are making good 
contact with the skin. If alarm is still on, detach the electrode and 
repeat this procedure with one of the other electrode lead wires 
supplied with your stimulator. If the alarm stops, the original lead 
wires may be defective. If the 

 icon appears, there is a problem 

with the stimulator. DO NOT ATTEMPT TO FIX IT YOURSELF. 
Contact Biomet or your sales representative. 

Helpful Tips

Loose Electrodes

 - Make sure that both electrodes are in complete 

contact with your skin. Moisten or replace worn electrodes if 
necessary.

Incomplete Circuit

 - Check all connection points, insuring tight fit 

of lead wire plug into the Biomet

®

 OrthoPak

®

 Non-invasive Bone 

Growth Stimulator System device and full engagement of electrode 
cable pin into electrode wire receptacle.

Broken Electrode Lead Wire

 - If you have checked the electrodes 

and the connections and the alarm continues, a break in the lead 
wire may be responsible. Remove the old electrode lead wire. 
Attach a new electrode lead wire into the jack. An extra electrode 
lead wire is provided in the assembly. If the alarms still continue, 
please call Biomet.

When the electrodes are 
applied and are making 
good contact with the 
skin, there will be a 
blinking indicator  “ ” 
icon (Figure 4).

4

Summary of Contents for OrthoPak

Page 1: ...Biomet OrthoPak Non invasive Bone Growth Stimulator System Patient Manual ...

Page 2: ...his procedure bone is taken from another location of the patient s body and surgically implanted at the nonunion site Metal rods or plates and bone screws also may be attached to the broken bone to secure the ends in place while it mends An alternative nonsurgical procedure was developed for the treatment of nonunions This method incorporated the transmission of an electrical signal through the no...

Page 3: ...ns Wearing the Biomet OrthoPak Non invasive Bone Growth Stimulator System The Biomet OrthoPak Non invasive Bone Growth Stimulator System has been designed so that it is convenient to use comfortable to wear and safe to operate You should begin using the Biomet OrthoPak Non invasive Bone Growth Stimulator System device immediately after you have read the directions for use Directions for Use It is ...

Page 4: ...pe Electrode Covers Covers are to be used for non casted applications by patients who are experiencing difficulty keeping the electrode pads in good contact with the skin When the electrode retainer is properly applied over the electrode pad you will be able to shower without removing the electrode pad Figure 2 Device Holster The device holster is designed to securely hold the Biomet OrthoPak Non ...

Page 5: ...vice it is recommended that you take the second battery pack place it into the charger and charge fully In the meantime you may use the first battery pack to begin your treatment immediately NOTE The first battery pack may not provide a 24 hour treatment initially Additional batteries are available through Biomet customer Service Each day preferably at the same time you should change the battery f...

Page 6: ...s unit in any other device Use of the Biomet OrthoPak Non invasive Bone Growth Stimulator System battery packs in any other device may cause damage or malfunction to the batteries and or devices Do not short circuit overcharge crush mutilate nail penetrate heat reverse the or terminals or disassemble the battery pack Do not allow metal objects to come into contact with the battery pack terminals T...

Page 7: ...er is appropriately plugged into the A C outlet Remove the current battery pack from the Biomet OrthoPak Non invasive Bone Growth Stimulator System and place that battery pack in the charger cradle for charging Open the battery compartment by pressing the battery cover release button 1 and sliding the battery cover to open 2 4 5 Step 1 Continued Battery Charging ...

Page 8: ...cide to place the stimulator controller in the holster place the holster in a comfortable and convenient location that is close to your fracture site Connect the two electrode wires into the electrode lead wire Insert the electrode lead wire plug into the jack at the top of the stimulator controller Figure 1 Choose the electrode lead wire that reaches from the location of the device or device hols...

Page 9: ... your sales representative Helpful Tips Loose Electrodes Make sure that both electrodes are in complete contact with your skin Moisten or replace worn electrodes if necessary Incomplete Circuit Check all connection points insuring tight fit of lead wire plug into the Biomet OrthoPak Non invasive Bone Growth Stimulator System device and full engagement of electrode cable pin into electrode wire rec...

Page 10: ...ck is already fully charged the charger will terminate the recharging process early This will be indicated by the charger having no orange light when charging is complete Do not be concerned if the battery pack is kept in the charger cradle for a long period of time When the Biomet OrthoPak Non invasive Bone Growth Stimulator System Needs a Freshly Charged Battery Pack the Following will Occur 1 T...

Page 11: ...r will not treat until USB cable is disconnected Contact Biomet Troubleshooting Electrodes Change your electrodes as required Different skin types will provide for a longer or shorter life of the electrodes If the alarm indicates a disconnection it is likely that either the lead wire connection is incomplete or the adhesive gel on the electrode is no longer working and the electrodes need changing...

Page 12: ...cords the number of days you receive treatment This helps your doctor track your treatment Helpful Tips Keep the audible alarm setting in the ON position as much as possible This alarm will help warn you of any problems with the device During special occasions when you would like the device not to tell you audibly about stimulator problems you may press the button for 3 seconds to turn off the aud...

Page 13: ...rly If You Have Questions If you have questions about your Biomet OrthoPak Non invasive Bone Growth Stimulator System contact BIOMET domestically at 1 800 526 2579 or 1 973 299 9300 if calling from outside the United States Representatives are available from 8 30am to 5 00pm EST Monday through Friday At other times please leave a clear message for a return call by the next business day IMPORTANT A...

Page 14: ...allows full weight bearing on the casted extremity unless gross motion greater than 5 degrees in any plane at the nonunion site is present In such case weight bearing is not advised and should not be permitted as this may compromise the effectiveness of the treatment The safety and effectiveness of the use of the device on individuals lacking skeletal maturity has not been established In the prese...

Page 15: ...ction generally resolves spontaneously following diagnosis and correction of the underlying cause Electrode Instructions for Use Do not open outer packet until ready to use 1 Tear open packet 2 Remove electrode from clear plastic backing liner 3 Wet finger with tap water and moisten entire gel area 4 Place electrode on skin 5 Connect electrode to electrode lead Renewal 1 With continuous use electr...

Page 16: ...ith EBI lineage Biomet has helped over one million people To learn more about this product contact your local Biomet Sales Representative today PN 1067800 00 Rev D Biomet OrthoPak Non invasive Bone Growth Stimulator System Patient Manual 399 Jefferson Road Parsippany NJ 07054 800 526 2579 www biomet com BNS231001L 03 14 2014 EBI LLC All trademarks are the property of Biomet Inc or one of its subsi...

Reviews: