bioMerieux MINI VIDAS User Manual Download Page 3

3

HOW TO PERFORM A CALIBRATION? 

Read the MLE bar code

 

Integrate factory data into the instrument

1

When using a new lot

Calibrate

 

Adjustment of the instrument

to the factory data

2

When using a new lot

+

Every 14 or 28 days

Control

 

Check calibration and ensure that reagent

performance has not been altered

After each calibration

+

When using a new kit

HOW TO SCAN MLE DATA?

 

In the

 MAIN MENU

, select the 

Master Lot Menu

.

Select 

Scan Master Lot

.

Scan the MLE barcode on the reagent kit label using the bar code reader :

slowly scan the barcode from top to bottom or from bottom to top until the code has 

been read completely.

At the end of the reading, 
“Please wait” is displayed.

 

The MLE card is then 
automatically printed. 

 

1 2 3

4 5 6
7 8 9

0 .

?

HOW TO RUN STANDARDS (S)

AND CONTROLS (C) FOR CALIBRATION?

In the 

MAIN MENU

, select 

Status Screen

.

THEN PRESS

Repeat this operation for each standard and control used for the test.
Load SPR

®

, Strips and distribute Standard and Controls in the strips.

1 2 3

4 5 6
7 8 9

0 .

?

1 2 3

4 5 6
7 8 9

0 .

?

1 2 3

4 5 6
7 8 9

0 .

?

2

.  Select the1st position

Typing 

1

.

3

.   Select 

S

 for the standard,

or

 C

 for the control.

4

.   Type 

for the standard

number 1…

Run the analysis selecting “

START

”.

1

. Select section 

A

 or 

B

.

?

?

Document only for Training use. Does not replace the User Manual.

Refer to MINI VIDAS

®

 user’s manual for detailed data