16 OF 34
|
OPERATING MANUAL
|
ULT25NEU
|
10.05.22
4.2.2
Change the Setpoint to a New Value
There are two ways to change the Setpoint.
>
Manually adjust the temperature using
^
/
v
.
>
Use one of the three adjustable
Preset Buttons
.
Adjusting the Setpoint manually.
>
Because of the importance of the Setpoint Temperature, changing
the Setpoint requires a “two-button procedure”. This means:
–
Method A
– press/hold the
Set Temp
button while simultaneously
pressing
^
/
v
.
OR
–
Method B
– press/hold the
Set Temp
button until the Set Temp LED
glows and then press
^
/
v
.
(Methods A and B achieve the same effect. The user can use
whichever method they prefer.)
To change the Setpoint by
Method A
>
Press and hold the
Set Temp
button and then push the
^
/
v
buttons
(do not release the
Set Temp
button).
–
The Setpoint temperature will appear with the prefix S in the LCD display.
–
When the
^
/
v
buttons are first pushed the Setpoint will change by
1ºC at first.
–
If the
^
/
v
buttons are held for five seconds the step size will increase/
decrease by 5ºC.
–
If the
^
/
v
buttons are briefly released (less than a half second) while
continuing to hold down the
Set Temp
button the increment/decrement
size will return to 1ºC.
–
Release all buttons once the desired Setpoint is achieved. The display
will return to chamber temperature after about eight seconds.