15
/
19
You
may
press
“M”
button
to
select
CTL
mode,
for
selecting
the
CTL
mode,
please
refer
to
the
“THE
TWO
MEASURING
MODES”
section.
Step
3
Apply
the
control
solution
Shake
the
control
solution
vial
thoroughly
before
use.
Squeeze
out
a
drop
and
wipe
it
off,
then
squeeze
another
drop
and
place
it
on
the
tip
of
the
vial
cap.
Hold
the
meter
to
move
the
absorbent
hole
of
test
strip
so
it
touches
the
drop.
Once
the
confirmation
window
fills
completely,
the
meter
will
begin
counting
down.
To
avoid
contaminating
the
control
solution,
do
not
directly
apply
control
solution
onto
a
strip.
Step
4
The
result
of
the
control
solutions
test
appears
on
the
display.
The
result
will
not
be
stored
under
CTL
mode.
The
result
of
the
control
solutions
test
should
be
within
the
range
printed
on
the
test
strip
vial
label.
If
the
test
results
falls
outside
the
specified
range
printed
on
test
strip
vial
label,
repeat
the
test.
Results
falling
outside
the
specified
range
may
be
caused
by:
■
Error
in
performing
the
test
■
Expired
or
contaminated
Control
Solution
■
Expired
or
contaminated
Test
Strip
■
Malfunc on
of
your
Meter
If
you
continue
to
get
control
solution
test
results
that
fall
outside
of
the
range
printed
on
the
vial,
the
system
may
not
be
working
properly,
Do
not
use
the
meter.
Please
contact
your
customer
support
immediately
NOTE:
Do
not
use
the
expired
control
solution.
VIEWING
RESULTS
ON
THE
METER
Your
meter
stores
the
500
most
recent
blood
glucose
test
results
with
dates
and
times.
It
also
provides
7
‐
,
14
‐
,
21
‐
,
28
‐
,
60
‐
and
90
‐
day
averages
of
your
blood
glucose
test
results.
Note:
The
control
solution
readings
are
not
included
in
the
average
of
your
results.
Recalling
the
Stored
Test
Results
1.Start
with
the
meter
off,
press
the
“M”
button
to
turn
on
the
meter.
The
“
”
symbol
will
flash
with
the
time
and
date
displayed
on
the
screen.
2.Press
“M”
again,
the
most
recent
test
results
with
date
and
time
will
display.
Press
“M”
once
more
and
the
next
most
recent
test
results
will
appear.
Each
time
you
press
and
release
“M”,
the
meter
will
recall
up
to
your
last
500
test
results
in
order.
Press
the
“S”
to
turn
off
the
meter.
When
the
memory
is
full,
the
oldest
result
is
dropped
and
the
newest
result
is